FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2513077
·
Received April 2, 2012
Report
- Report Number
- 2531779-2012-02670
- Event Type
- Malfunction
- Date Received
- April 2, 2012
- Report Date
- January 18, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON 03/03/2012 WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. CONTAMINATION WAS OBSERVED ON THE FORCE SENSOR ASSEMBLY. UNRELATED TO THE EVENT THE DISPLAY WAS FOUND TO HAVE A WHITE SPOT. CORRECTION/REMOVAL REPORT NUMBER - 2531779-03/24/2010-003-R.
Description of Event or Problem · 1
THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON 03/03/2012 WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. CONTAMINATION WAS OBSERVED ON THE FORCE SENSOR ASSEMBLY. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 YR |