FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION¿ S8 SYSTEM

MDR report key: 25125522 · Received May 8, 2026

Report

Report Number
1723170-2026-00724
Event Type
Malfunction
Date Received
May 8, 2026
Date of Event
May 5, 2026
Report Date
May 8, 2026
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
HAW
UDI-DI
00763000359881
PMA / PMN Number
K162309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9735836, SERIAL/LOT #: -, UBD: , UDI#: H3, H6: A MANUFACTURER REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. IN SUMMARY, A SYSTEM CHECKOUT WAS PERFORMED, HARDWARE WAS REPLACED, AND THE SYSTEM PERFORMED AS INTENDED. CODES FDM B01, FDR C04, FDC D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: THE CAMERA ETHERNET CABLE (PRODUCT: 9735836, LOT: 211013) WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE CAMERA ETHERNET CABLE WAS FOUND TO BE FULLY FUNCTIONAL WITH NO PROBLEM FOUND. THE REPORTED EVENT COULD NOT BE DUPLICATED BY MEDTRONIC PERSONNEL. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE CABLE COMING OUT OF THE ARM CONNECTING INTO THE CAMERA WAS FRAYED AND THE INTERNAL WIRES WERE EXPOSED. FUNCTIONALITY WAS NOT AFFECTED. THERE WAS NO PATIENT INVOLVEMENT.