FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE

MDR report key: 25116661 · Received May 7, 2026

Report

Report Number
9610773-2026-03614
Event Type
Malfunction
Date Received
May 7, 2026
Report Date
May 7, 2026
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761075015
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. BASED ON THE RESULTS OF THE INVESTIGATION, A PROBABLE ROOT CAUSE COULD NOT BE IDENTIFIED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE CAUSE OF THE DEFORMED OUTER TUBE WAS TRACED TO COMPONENT FAILURE, HOWEVER THE CAUSE OF THE FOREIGN MATERIAL FOUND IN THE LLK PLUG (LASER LIGHT CABLE) OF THE VIDEO CABLE SECTION COULD NOT BE ESTABLISHED. DATE OF EVENT IS UNKNOWN. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE HAD FOREIGN MATERIAL IN THE LLK (LASER LIGHT CABLE) PLUG OF THE VIDEO CONNECTOR AND EXHIBITED DEFORMED OUTER TUBE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
118122 VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE RIGID VIDEO LAPAROSCOPE HET OLYMPUS WINTER & IBE GMBH WA50052A 04042761075015

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown