BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES
Report
- Report Number
- 9617032-2026-00808
- Event Type
- Malfunction
- Date Received
- May 7, 2026
- Date of Event
- March 24, 2026
- Report Date
- May 11, 2026
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- UDI-DI
- 30382903679554
- PMA / PMN Number
- BK050036
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY BD DID NOT RECEIVE ANY SAMPLES OR PHOTOS FOR INVESTIGATION. HOWEVER, 20 RETAINED SAMPLES FROM THE REPORTED LOT WERE UTILIZED FOR FUNCTIONAL TESTS. ALL 20 SAMPLES WERE FOUND TO BE WITHIN SPECIFICATION. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE LOT 5162593, FOR THE INDICATED FAILURE MODE: DRAW VOLUME. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
REPORT 2 OF 2. IT WAS REPORTED WHILE USING BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
REPORT 2 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 240230 | BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES | TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION | JKA | BECTON, DICKINSON AND COMPANY (BD) | 5162593 | 30382903679554 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |