ALINITY I PROCESSING MODULE
Report
- Report Number
- 3016438761-2026-00260
- Event Type
- Malfunction
- Date Received
- May 7, 2026
- Date of Event
- April 14, 2026
- Report Date
- May 14, 2026
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- JJE
- UDI-DI
- 00380740137366
- PMA / PMN Number
- K170317
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER REPORTED FALSE REACTIVE ALINITY I HAVAB IGG II RESULTS FOR MULTIPLE PATIENTS. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2026, SID (B)(6) (19-YEAR-OLD MALE): INITIAL RESULT ON (B)(6) = POSITIVE, REPEAT ON ANOTHER ALINITY = NEGATIVE. (B)(6) 2026, SID (B)(6) (33-YEAR-OLD MALE): INITIAL RESULT ON (B)(6) = POSITIVE, REPEAT = NEGATIVE, REPEAT ON ANOTHER ALINITY = NEGATIVE. ADDITIONALLY, THE CUSTOMER NOTED A SHIFT UP IN QUALITY CONTROLS. A PRECISION STUDY WAS COMPLETED WITH SID (B)(6) FOR TROUBLESHOOTING PURPOSES: (B)(6) = 0.3, (B)(6) = 0.75, (B)(6) = 0.43, (B)(6) = 1.32, (B)(6) = 0.53, (B)(6) = 1.09, (B)(6) = 0.28, (B)(6) = 1.07, (B)(6) = 0.41, (B)(6) = 0.4. A PATIENT COMPARISON STUDY WAS COMPLETED FOR TROUBLESHOOTING PURPOSES: SAMPLE 1: (B)(6) = 10.8, (B)(6) = 9.61, SAMPLE 2: (B)(6) = 0.14, (B)(6) = 0.12, SAMPLE 3: (B)(6) = 7.25, (B)(6) = 7.48, SAMPLE 4: (B)(6) = 9.08, (B)(6) = 8.53, SAMPLE 5: (B)(6) = 8.07, (B)(6) = 8.18, SAMPLE 6: (B)(6) = 11.03, (B)(6) = 10.20, SAMPLE 7: (B)(6) = 1.26, (B)(6) = 0.73, SAMPLE 8: (B)(6) = 0.34, (B)(6) = 0.34, SAMPLE 9: (B)(6) = 0.09, (B)(6) = 0.10, SAMPLE 10: (B)(6) = 0.21, (B)(6) = 0.27. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 213312 | ALINITY I PROCESSING MODULE | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | ABBOTT LABORATORIES | 00380740137366 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |