I-STAT TROPONIN (HS-TNI ) CARTRIDGE
Report
- Report Number
- 2245578-2026-00097
- Event Type
- Malfunction
- Date Received
- May 7, 2026
- Date of Event
- April 9, 2026
- Report Date
- May 7, 2026
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- MMI
- UDI-DI
- 00054749003051
- PMA / PMN Number
- K240984
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT#: (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON 21-APR-2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT FALSE POSITIVE RESULT ON A 70-YEAR-OLD FEMALE PATIENT WITH BURNING PAIN TO THE ENTIRE BACK THEN BOTH ARMS. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: DATE: TESTED: RESULTS: SAMPLE: POSITIVE/NEGATIVE: I-STAT, ON (B)(6) 2026, 19:00 >1000 NG/L, A, POSITIVE, ROCHE, ON (B)(6) 2026, 23:26 <6 NG/L, B, NEGATIVE, ROCHE, ON (B)(6) 2026, 02:30 <6 NG/L, C, NEGATIVE. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY. 99TH PERCENTILE: 90% CI: SEX: N, (NG/L, PG/ML), (NG/L, PG/ML), FEMALE: 490, 13, (10, 17), MALE: 404, 28, (19, 58), OVERALL: 895, 21, (14, 30).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 433729 | I-STAT TROPONIN (HS-TNI ) CARTRIDGE | HS-TNI CARTRIDGE | MMI | ABBOTT POINT OF CARE INC. | NA | B26028 | 00054749003051 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |