IMP,TSV,3.7,10,MTX,MG
Report
- Report Number
- 0001038806-2026-02480
- Event Type
- Injury
- Date Received
- May 7, 2026
- Date of Event
- April 13, 2026
- Report Date
- May 7, 2026
- Manufacturer
- ZIMVIE US CORP LLC
- Product Code
- DZE
- UDI-DI
- 00889024019928
- PMA / PMN Number
- K101977
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). A1: PATIENT IDENTIFIER: UNKNOWN / NOT PROVIDED. A4: PATIENT WEIGHT: UNKNOWN / NOT PROVIDED. D10: CONCOMITANT MEDICAL PRODUCT: TSVTB10, IMP,TSV,3.7,10,MTX,MG, LOT: 1296927. TSVT4B13, IMP,TSV,4.1,13,MTX,MG, LOT: 1252054. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT 3 IMPLANTS AT TOOTH SITES #23, 25 AND 26 WERE REMOVED DUE TO A BONE LOSS. OSSEOINTEGRATED IMPLANTS FOR 6 MONTHS THAT HAVE COMPLETELY RESORBED THE BUCCAL BONE PLATE. PROCEDURE WAS COMPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 294042 | IMP,TSV,3.7,10,MTX,MG | DENTAL IMPLANT | DZE | ZIMVIE US CORP LLC | 1296927 | 00889024019928 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |