BD VACUTAINER® K2 EDTA (K2E) 5.4MG BLOOD COLLECTION TUBES
Report
- Report Number
- 1917413-2026-00403
- Event Type
- Malfunction
- Date Received
- May 6, 2026
- Date of Event
- April 10, 2026
- Report Date
- April 14, 2026
- Manufacturer
- BECTON DICKINSON & CO (FRANKLIN LAKES)
- Product Code
- GIM
- UDI-DI
- 30382903678564
- PMA / PMN Number
- BK230980
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THERE WERE MULTIPLE MEDICAL DEVICE TYPES REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL DEVICE TYPE IS AS FOLLOWS. D.2.B MEDICAL DEVICE TYPE: JKA. D.2.A COMMON DEVICE NAME: TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K213670; K231373. INVESTIGATION SUMMARY: BD RECEIVED 1 PHOTO FOR INVESTIGATION. EVALUATION OF THE PHOTO SHOWED THE PRESENCE OF FOREIGN MATTER IN THE TUBE. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE LOTS 5260171 AND 5260179, FOR THE INDICATED FAILURE MODE: FOREIGN MATTER. NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.
REPORT 1 OF 2: IT WAS REPORTED PRIOR TO USING BD VACUTAINER® K2 EDTA (K2E) 5.4MG BLOOD COLLECTION TUBES, FOREIGN MATTER WAS FOUND IN 152 TUBES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 456819 | BD VACUTAINER® K2 EDTA (K2E) 5.4MG BLOOD COLLECTION TUBES | TUBES, VACUUM SAMPLE, WITH ANTICOAGULANT | GIM | BECTON DICKINSON & CO (FRANKLIN LAKES) | 5260171 | 30382903678564 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |