FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® K2 EDTA (K2E) 5.4MG BLOOD COLLECTION TUBES

MDR report key: 25107166 · Received May 6, 2026

Report

Report Number
1917413-2026-00403
Event Type
Malfunction
Date Received
May 6, 2026
Date of Event
April 10, 2026
Report Date
April 14, 2026
Manufacturer
BECTON DICKINSON & CO (FRANKLIN LAKES)
Product Code
GIM
UDI-DI
30382903678564
PMA / PMN Number
BK230980
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE MEDICAL DEVICE TYPES REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL DEVICE TYPE IS AS FOLLOWS. D.2.B MEDICAL DEVICE TYPE: JKA. D.2.A COMMON DEVICE NAME: TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K213670; K231373. INVESTIGATION SUMMARY: BD RECEIVED 1 PHOTO FOR INVESTIGATION. EVALUATION OF THE PHOTO SHOWED THE PRESENCE OF FOREIGN MATTER IN THE TUBE. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE LOTS 5260171 AND 5260179, FOR THE INDICATED FAILURE MODE: FOREIGN MATTER. NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

REPORT 1 OF 2: IT WAS REPORTED PRIOR TO USING BD VACUTAINER® K2 EDTA (K2E) 5.4MG BLOOD COLLECTION TUBES, FOREIGN MATTER WAS FOUND IN 152 TUBES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
456819 BD VACUTAINER® K2 EDTA (K2E) 5.4MG BLOOD COLLECTION TUBES TUBES, VACUUM SAMPLE, WITH ANTICOAGULANT GIM BECTON DICKINSON & CO (FRANKLIN LAKES) 5260171 30382903678564

Patients

Seq Age Sex Outcome Treatment
1