ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2012-02608
- Event Type
- Malfunction
- Date Received
- March 29, 2012
- Date of Event
- February 29, 2012
- Report Date
- March 2, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(6). THE CARTRIDGES HAVE NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
(B)(4): THE CARTRIDGE WAS NOT RETURNED FOR INVESTIGATION. A RETAIN SAMPLE CARTRIDGE FROM LOT # B201710 HAS BEEN EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST AND A FILL TEST WAS PERFORMED WITH NO FAILURES OBSERVED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
THE DISTRIBUTOR CONTACTED ANIMAS ON MARCH 2 AND REPORTED THAT THERE WERE AIR BUBBLES IN THE CARTRIDGE AND TUBING. THERE WERE NO LEAKS IDENTIFIED IN THE CARTRIDGE, AT THE TUBING, OR AT THE SKIN SITE. THE LUER LOCK WAS SECURE AND THE PATIENT WAS FILLING THE CARTRIDGE WITH ROOM TEMPERATURE INSULIN. THE PATIENT WAS ALSO REPORTEDLY PRESSURIZING THE VIAL OF INSULIN SLOWLY AND ACCORDING TO THE INSTRUCTIONS FOR USE. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THERE WERE AIR BUBBLES REPORTEDLY FORMED IN THE CARTRIDGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM IR1200/1250/2020/OTP CART |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |