FDA Adverse Event Malfunction Summary report: N

ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT

MDR report key: 25101109 · Received May 6, 2026

Report

Report Number
3005094123-2026-00212
Event Type
Malfunction
Date Received
May 6, 2026
Date of Event
April 18, 2026
Report Date
May 6, 2026
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MMI
UDI-DI
00380740160647
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 08P13, THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 04Z21, WITH 510K/PMA/BLA NUMBER K202525.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULT GENERATED ON THE ALINITY I PROCESSING MODULE (SN: (B)(6) FOR A MALE PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE: <34.2 NG/L FOR MALE PATIENTS): (B)(6) 2026 INITIAL RESULT FOR SAMPLE ID (B)(6) = 608.1 NG/L. THIS RESULT WAS QUESTIONED. THE SAME SAMPLE WAS REPEATED ON ANOTHER ALINITY INSTRUMENT (SN: (B)(6), AND THE RESULT = 12.9 NG/L A SECOND SAMPLE WAS OBTAINED FROM THE SAME PATIENT. SAMPLE ID (B)(6) RESULT ON INSTRUMENT (B)(6) = 16.8 NG/L, REPEAT RESULT ON INSTRUMENT (B)(6) = 14.5 NG/L. A THIRD SAMPLE WAS OBTAINED FROM THE SAME PATIENT (SAMPLE ID (B)(6) AND THE RESULT ON INSTRUMENT (B)(6) = 20.1 NG/L, RESULT ON INSTRUMENT (B)(6) = 17.2 NG/L. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
587246 ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ABBOTT IRELAND DIAGNOSTICS DIVISION 82211UD00 00380740160647

Patients

Seq Age Sex Outcome Treatment
1