FDA Adverse Event Injury Summary report: N

BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE

MDR report key: 25095818 · Received May 6, 2026

Report

Report Number
1024879-2026-00721
Event Type
Injury
Date Received
May 6, 2026
Date of Event
March 5, 2026
Report Date
April 21, 2026
Manufacturer
BECTON DICKINSON & CO (FRANKLIN LAKES)
Product Code
JKA
UDI-DI
30382903686088
PMA / PMN Number
K243207
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H4. DEVICE MANUFACTURE DATE: UNKNOWN. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K982541. INVESTIGATION SUMMARY: BD RECEIVED TWO PHOTOS FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR DEFECTIVE LOCKING MECHANISM WAS OBSERVED. BD WAS UNABLE TO DETERMINE THE SPECIFIC LOT NUMBER ASSOCIATED WITH THIS COMPLAINT; THEREFORE, A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE CONDUCTED. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE DEFECTIVE LOCKING MECHANISM. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

REPORT 1 OF 3. IT WAS REPORTED THAT WHEN THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WAS IN PROCESS OF BEING REMOVED, THE SAFETY WAS UNINTENTIONALLY KNOCKED OUT OF PLACE BY THE PATIENT AND STAFF COULD NOT ACTIVATE THE SAFETY AND WAS STUCK BY THE NEEDLE. STAFF MEMBER FOLLOWED STANDARD NEEDLESTICK PROTOCOL PER FACILITY. IT IS UNSPECIFIED IF MEDICAL INTERVENTION OR TREATMENT TOOK PLACE.