FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET

MDR report key: 25095040 · Received May 6, 2026

Report

Report Number
1024879-2026-00718
Event Type
Malfunction
Date Received
May 6, 2026
Date of Event
April 8, 2026
Report Date
May 11, 2026
Manufacturer
BECTON DICKINSON & CO (FRANKLIN LAKES)
Product Code
JKA
UDI-DI
30382903686538
PMA / PMN Number
K980414
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D.4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN H.4. DEVICE MANUFACTURE DATE: UNKNOWN E.1. INITIAL REPORTER FACILITY NAME: (B)(6). THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K252506 A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY BD DID NOT RECEIVE ANY SAMPLES OR PHOTOS FOR INVESTIGATION. TESTING OF RETAINED SAMPLES COULD NOT BE PERFORMED AS THE BATCH NUMBER IS UNKNOWN. A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE COMPLETED AS THE BATCH NUMBER IS UNKNOWN. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE INDICATED FAILURE MODES: DIFFICULT/UNABLE TO OPERATE AND NEEDLE POINT INTEGRITY. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET, THE SAFETY MECHANISM OF AN UNSPECIFIED NUMBER OF DEVICES WAS HARD TO ACCESS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET, THE SAFETY MECHANISM OF AN UNSPECIFIED NUMBER OF DEVICES WAS HARD TO ACCESS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
121051 BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION JKA BECTON DICKINSON & CO (FRANKLIN LAKES) UNKNOWN 30382903686538

Patients

Seq Age Sex Outcome Treatment
1