FDA Adverse Event Malfunction Summary report: N

2.25MM PEDICLE PROBE

MDR report key: 25087830 · Received May 5, 2026

Report

Report Number
1723170-2026-00700
Event Type
Malfunction
Date Received
May 5, 2026
Date of Event
March 16, 2026
Report Date
May 5, 2026
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
HAW
UDI-DI
00763000504304
PMA / PMN Number
K050438
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

H3: THE 2.25MM PEDICLE PROBE, LOT NUMBER: 191107, WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE RETURNED PROBE HAD A SEVERELY BENT TIP. OTHERWISE, WITH MARKERS ATTACHED AND FULLY SEATED, THE PROBE DISPLAYED GOOD GEOMETRY AND DIVOT ERROR READINGS WITH NORMAL TRACKING AND VERIFICATION. H6: CODES B01, C07, AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT TIP DEFORMATION WAS CONFIRMED DURING THE ACCEPTANCE INSPECTION. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
285111 2.25MM PEDICLE PROBE NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC. 9733457 191107 00763000504304

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown