CUBBY BEDS
Report
- Report Number
- 3016541541-2026-00035
- Event Type
- Malfunction
- Date Received
- May 5, 2026
- Date of Event
- April 7, 2026
- Report Date
- May 5, 2026
- Manufacturer
- SENSORY MEDICAL INC.
- Product Code
- OYS
- UDI-DI
- 00860007550249
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
SENSORY MEDICAL REQUESTED RETURN FOR EXAMINATION OF THE DAMAGED SAFETY SHEET. THE PRODUCT HAS NOT BEEN RETURNED FOR EVALUATION AS OF THE WRITING OF THIS INVESTIGATION. MULTIPLE STATEMENTS ARE MADE IN THE CUBBY BED USER MANUAL REGARDING SAFE USE OF THE BED TO PREVENT ELOPEMENT AND OTHER SAFETY CONCERNS. PAGE 3 OF THE CUBBY BED USER MANUAL CONTAINS A WARNING FOR ENTRAPMENT HAZARD. "TO AVOID DANGEROUS GAPS DO NOT USE THIS PRODUCT WITHOUT THE SAFETY SHEETS COMPLETELY ZIPPED AND LOCKED IN PLACE TO THE SIDEWALL. SAFETY SHEET WITH LOCK IN PLACE IS ALWAYS REQUIRED." PAGE 3 OF THE CUBBY BED USER MANUAL CONTAINS A WARNING FOR "USE THE MAINTENANCE CHECKLIST IN THE MANUAL TO INSPECT THE CANOPY, SEAMS, ZIPPERS, ELECTRONIC ACCESSORIES, CORDS, METAL FRAME, SCREWS, SLATS, LOCKS, AND SAFETY SHEETS EVERY 30 DAYS." PAGE 3 OF THE CUBBY BED USER MANUAL CONTAINS A WARNING FOR "IF ANY DAMAGE IS PRESENT, IMMEDIATELY DISCONTINUE USE AND CONTACT CUBBY FOR REPAIR OR REPLACEMENT." PAGE 5 CONTAINS A PARTS LIST, WHICH INCLUDES THE SAFETY SHEETS AND LOCKS. PAGE 11 INSTRUCTS TO SECURE THE SAFETY SHEET ZIPPER TO THE LOOP ON THE CANOPY WITH THE LOCK TO PREVENT THE PATIENT USER FROM REMOVING THE SHEET. DO NOT USE THE PRODUCT WITHOUT THE SAFETY SHEETS ZIPPED AND LOCKED IN PLACE. ON PAGE 15 "WHAT HAPPENS IF MY CUBBY RIPS OR I NOTICE OTHER DAMAGE? PLEASE CONTACT US RIGHT AWAY. ONLY USE YOUR CUBBY BED WHEN IT IS ASSEMBLED CORRECTLY AND IS IN SAFE WORKING CONDITION." PAGE 15 ASSEMBLY + CARE FAQ'S INCLUDES DESCRIPTION AND USE OF THE LOCKS ON THE SAFETY SHEETS AND THE TECHNOLOGY HUB ZIPPERS. PAGE 22 INCLUDES A MONTHLY MAINTENANCE CHECKLIST, ONE CHECK IS: "SAFETY SHEET FABRIC, CORD LOOP AND ZIPPERS HAVE NO TEARS, RIPS OR SNAGS AND LOCK IS SECURED." ON PAGE 22 INCLUDES A MONTHLY MAINTENANCE CHECKLIST, ONE CHECK IS: "EACH ZIPPER ON THE CANOPY IS FUNCTIONAL AND SEAMS OF ZIPPERS ARE INTACT." PAGE 22 OF THE USER MANUAL, MAINTENANCE CHECKLIST, DESCRIBES DISCONTINUING USE OF THE BED IF ANYTHING IS DAMAGED OR MISSING. ADDITIONAL INVESTIGATION IS NEEDED, AND SENSORY MEDICAL'S ROOT CAUSE INVESTIGATION IS ONGOING. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
A COMPLAINT WAS REPORTED TO SENSORY MEDICAL BY A DURABLE MEDICAL EQUIPMENT (DME) REPRESENTATIVE. PER REPRESENTATIVE, THE CHILD BROKE THE SAFETY SHEET ZIPPER AND ELOPED THROUGH IT. PER REPRESENTATIVE, THE MOM SAID THE LOCK WAS INSTALLED AT THE TIME OF THE EVENT AND THE CHILD PULLED HARD ENOUGH TO BREAK THE ZIPPER WITH THE LOCK STILL ATTACHED. A PICTURE PROVIDED SHOWS THE UNINSTALLED SAFETY SHEET WITH A MISSING ZIPPER SLIDER, AND ANOTHER SHEET WITH A DAMAGED ZIPPER WITH FRAYING THREADS. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 498712 | CUBBY BEDS | CANOPY BED | OYS | SENSORY MEDICAL INC. | 231023M01 | 00860007550249 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male |