INSET GUARD
Report
- Report Number
- 8021545-2026-06018
- Event Type
- Malfunction
- Date Received
- May 5, 2026
- Date of Event
- December 4, 2025
- Report Date
- April 23, 2026
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- PMA / PMN Number
- K210544
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
E1: (B)(6). NAME: MEDTRONIC MINIMED COUNTRY: UNITED STATES OF AMERICA STREET: 18000 DEVONSHIRE STREET CITY: NORTHRIDGE STATE, PROVINCE OR TERRITORY: CALIFORNIA POST OFFICE OR ZIP CODE: 91325. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS H11: INVESTIGATION SUMMARY COMPLAINT INVESTIGATION RESULTS: REVIEW OF COMPLAINT RECORD DATABASE (B)(4) EVENT DESCRIPTION AND ALL AVAILABLE INFORMATION AND ASSIGNED MALFUNCTION CODE CONNECTION/ADHESIVE ISSUES- ACCIDENTAL - NOT INFUSION SET RELATED IT HAS BEEN DETERMINED THE COMPLAINT DOES NOT ALLEGE A PRODUCT MALFUNCTION HAS OCCURRED. THEREFORE, NO FURTHER INVESTIGATION IS REQUIRED. CONCLUSION OF COMPLAINT INVESTIGATION: LOT NUMBER 6012759 WAS PROVIDED, HOWEVER, AS NO PRODUCT MALFUNCTION WAS ALLEGED: NO VISUAL INSPECTION IS REQUIRED TO BE PERFORMED. NO RETAIN SAMPLE TESTING IS REQUIRED TO BE CONDUCTED. NO FURTHER ASSESSMENT WILL BE MADE REGARDING POTENTIAL PRODUCT PERFORMANCE ISSUES, COMPONENT INTEGRITY, OR MANUFACTURING DEFECTS. CORRECTIVE AND PREVENTIVE ACTION (CAPA) DETERMINATION: BASED ON THE AVAILABLE INFORMATION, NO FURTHER INVESTIGATION IS REQUIRED NOR CAPA PLAN IS NEEDED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH TRACKING AND TRENDING (MONTHLY TRIPS AND ALERTS). SUMMARY CONCLUSION: BASED ON THE AVAILABLE INFORMATION, THE INVESTIGATION HAS BEEN PERFORMED, AND THE COMPLAINT COULD NOT BE CONFIRMED AS ANALYZED. THEREFORE, THE COMPLAINT IS CLOSED BASED ON THE EVENT DESCRIPTION AND ALL INFORMATION MADE AVAILABLE. BASED ON THE AVAILABLE INFORMATION, THIS EVENT IS DEEMED TO BE A REPORTABLE MALFUNCTION. TO DATE NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. FDA REGISTRATION NUMBER REPORTING SITE: 8021545 MANUFACTURING SITE: 8021545.
IT WAS REPORTED THAT THE INFUSION ST TUBING DETACHED ON (B)(6) 2025 DUE TO INFUSION SET TUBING WAS CAUGHT IN DOORKNOB.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 555595 | INSET GUARD | SET, ADMINISTRATION, INTRAVASCULAR | FPA | UNOMEDICAL A/S | MMT-431AG600 | 6012759 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |