FDA Adverse Event Other Summary report: N

TRANSPORT CHAIR

MDR report key: 2507532 · Received March 22, 2012

Report

Report Number
1417592-2012-00013
Event Type
Other
Date Received
March 22, 2012
Date of Event
February 24, 2012
Report Date
March 20, 2012
Manufacturer
MEDLINE INDUSTRIES, INC.
Product Code
INM
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE END USER ATTEMPTED TO SIT DOWN ON THE CHAIR AND IT ROLLED BACK. IT IS NOT KNOW IF SHE ACTUALLY FELL OR WAS CAUGHT BY HER CAREGIVER. SHE WAS EVALUATED IN THE ER FOR CHEST PAIN. A T11 FRACTURE WAS IDENTIFIED VIA CHEST X-RAY. SHE HAS A HISTORY OF OSTEOPOROSIS AND OTHER SPINAL FRACTURES. THE SON STATED HE DID NOT KNOW IF THE T11 FRACTURE RESULTED FROM THIS INCIDENT. SHE RECEIVED A STEROID INJECTION FOR BACK PAIN. NO OTHER TREATMENT WAS INITIATED. THE SON REPORTED THERE HAD BEEN SPORADIC BRAKE ISSUES DURING THE PREVIOUS MONTHS BUT THEY CONTINUED TO USE THE DEVICE . HE ALSO NOTED THAT A NUT WAS MISSING FROM THE LEFT BRAKE LEVER. THE SAMPLE WAS RETURNED AND EVALUATED. THE SEAT SUPPORT TUBES WERE BENT AND THE NUT THAT HOLDS THE TOP PIVOTING BRAKE LEVEL WAS MISSING AND AS A RESULT, THE BRAKE HANDLE WAS BENT. WEAR ON THE HANDLE INDICATES THAT THE NUT WAS IN PLACE AT ONE TIME. IT IS NOT KNOWN WHEN IT CAME OFF. THE MISSING NUT MAKES THE BRAKE MORE DIFFICULT TO LOCK, ALTHOUGH ONCE LOCKED, IT WAS FULLY FUNCTIONAL AND PREVENTED THE ROTATION OF THE WHEEL. THE SAMPLE EVALUATION DID NOT IDENTIFY ANY MFG DEFECT. THE OVERALL WEAR OF THE CHAIR SUGGESTS IT WAS NOT MAINTAINED WHICH IS THE LIKELY ROOT CAUSE OF THE DAMAGES TO THE DEVICE AND A CONTRIBUTING FACTOR TO THE REPORTED INCIDENT.

Description of Event or Problem · 1

THE CHAIR ROLL BACKWARDS AS THE END USER WENT TO SIT DOWN AND SHE FELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRANSPORT CHAIR INM MEDLINE INDUSTRIES, INC. MDS808200W

Patients

Seq Age Sex Outcome Treatment
1 71 YR Other