FDA Adverse Event Other Summary report: N

RAPID RESPONSE HUT TABLE

MDR report key: 2507465 · Received March 2, 2012

Report

Report Number
1932056-2012-00001
Event Type
Other
Date Received
March 2, 2012
Date of Event
February 28, 2012
Report Date
February 29, 2012
Manufacturer
MEDICAL POSITIONING, INC.
Product Code
LGX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION UNCOVERED THAT THE USER WAS ATTEMPTING TO PLUG THE PRODUCT INTO THE 120 VAC UTILITY POWER WHILE THE POWER SWITCH IS IN THE "ON" POSITION. BECAUSE THERE IS A TRANSFORMER IN THE UNIT, THERE IS A MOMENTARY SPIKE IN THE CURRENT DRAW AND THEN LEVELS OUT. THE COMPANY ADVISED THE CUSTOMER TO MAKE SURE THAT THE PRODUCT'S POWER SWITCH IS IN THE "OFF" POSITION BEFORE CONNECTING IT TO UTILITY POWER. THE COMPANY ALSO OPENED CAPA (B)(4). UPON REVIEW OF PRODUCT LABELING IT WAS DISCOVERED THAT INSTRUCTIONS FOR PROPERLY CONNECTING THE PRODUCT TO UTILITY POWER ARE MISSING FROM THE OWNER'S MANUAL. THE COMPANY WILL UPDATE THE MANUAL WITH THE PROPER INSTRUCTIONS/WARNING FOR CONNECTING THE UNIT TO UTILITY POWER. THE COMPANY WILL SEND THE MANUAL UPDATE TO ALL NINE (9) CUSTOMERS WHO HAVE PURCHASED THE PRODUCT.

Description of Event or Problem · 1

AS THE NURSE PLUGGED THE PRODUCT INTO 120 VAC UTILITY POWER, THERE WAS A MOMENTARY SPARK BETWEEN THE OUTLET AND THE PLUG AND THE NURSE RECEIVED A MINOR SHOCK, BUT NO INJURY FROM THE SPARK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RAPID RESPONSE HUT TABLE RAPID RESPONSE HUT TABLE LGX MEDICAL POSITIONING, INC. 1217

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other