FDA Adverse Event Malfunction Summary report: N

BD GASPAK¿ EZ CAMPY CONTAINER SYSTEM SACHETS

MDR report key: 25071616 · Received May 4, 2026

Report

Report Number
2647876-2026-00048
Event Type
Malfunction
Date Received
May 4, 2026
Date of Event
April 9, 2026
Report Date
April 29, 2026
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
KZJ
UDI-DI
30382902606803
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H4. DEVICE MANUFACTURE DATE: UNKNOWN. D4. MEDICAL DEVICE LOT#: UNKNOWN. E1. ADDRESS INFORMATION WAS NOT ABLE TO BE OBTAINED; THEREFORE, MD WAS USED AS A PLACE HOLDER. INVESTIGATION SUMMARY: A COMPLAINT INVESTIGATION WAS PERFORMED, DUE TO CAMPY ISSUE SHOWING FALSE POSITIVE WITH GASPAK EZ CAMPY CONTAINER CATALOG 260680. THE INVESTIGATION REQUIRED TO TEST PRODUCT FOR FUNCTIONALITY (MICROBIOLOGICAL PERFORMANCE, GAS CONCENTRATION TEST, CONDENSATE TIME), VISUAL INSPECTION, INCOMING AND BATCH RECORD REVIEW. RETENTION SAMPLES UNABLE TO BE TESTED. BATCH NUMBER WAS NOT RECEIVED. RETURNED GOODS NOT RECEIVED FOR INVESTIGATION. INCOMING AND BATCH RECORD REVIEW UNABLE TO BE PERFORMED SINCE BATCH NUMBER WAS NOT PROVIDED BY CUSTOMER. FROM A PERIOD OF 1 YEAR (APR2025 TO 29APR2026) QUALITY NOTIFICATION (QN) SYSTEM WAS EVALUATED. NO QNS WERE ISSUED DUE TO PRODUCT PERFORMANCE FAILURE. A COMPLAINT HISTORY REVIEW WAS CONDUCTED. NO TREND WAS FOUND RELATING TO THE CLAIM REPORTED. COMPLAINT IS NOT CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD GASPAK¿ EZ CAMPY CONTAINER SYSTEM SACHETS, A FALSE POSITIVE RESULT WAS OBTAINED ON UNSPECIFIED NUMBER OF PATIENTS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470827 BD GASPAK¿ EZ CAMPY CONTAINER SYSTEM SACHETS DEVICE, GAS GENERATING KZJ BECTON DICKINSON CARIBE LTD. UNKNOWN 30382902606803

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown