FDA Adverse Event Malfunction Summary report: N

SYSTEM 6 SAGITTAL SAW

MDR report key: 2507095 · Received March 23, 2012

Report

Report Number
1811755-2012-01060
Event Type
Malfunction
Date Received
March 23, 2012
Date of Event
February 29, 2012
Report Date
March 1, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE WAS RECEIVED AT THE MFR FOR EVAL, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH THE BEARINGS, WHICH WERE REPLACED ALONG WITH OTHER COMPONENTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT HANDPIECE OVERHEATED WHILE THE EQUIPMENT WAS BEING TESTED. THERE WAS NO PT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYSTEM 6 SAGITTAL SAW KIJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK