FDA Adverse Event Injury Summary report: N

PRIMA HUMERAL SYSTEM

MDR report key: 25067086 · Received May 4, 2026

Report

Report Number
3008021110-2026-00197
Event Type
Injury
Date Received
May 4, 2026
Date of Event
March 10, 2026
Report Date
May 4, 2026
Manufacturer
LIMACORPORATE S.P.A
Product Code
PHX
UDI-DI
08033390202954
PMA / PMN Number
K212800
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

REVIEW OF MANUFACTURING RECORDS DID NOT IDENTIFY ANY PRE-EXISTING ANOMALY OR NON CONFORMITY THAT COULD HAVE CONTRIBUTED TO REPORTED ISSUE. FROM FOLLOW-UP COMMUNICATION THE COMPLAINT SOURCE CONFIRMED THAT SURGEON TESTED THE STABILITY OF THE ASSEMBLY DURING ORIGINAL SURGERY, WITH NO ISSUES REPORTED. MOREOVER, THE EXPLANTED COMPONENTS WERE REPORTED TO BE IN GOOD CONDITION WHEN EXPLANTED, THUS EXCLUDING AN EARLY FAILURE OF THE COMPONENTS CAUSING THE ISSUE. THE MANUFACTURER WILL SUBMIT A FINAL REPORT AS SOON AS THE INVESTIGATION IS COMPLETED.

Description of Event or Problem · 0

REVISION SURGERY WAS PERFORMED ON (B)(6) 2026 DUE TO SHOULDER DISLOCATION. ORIGINAL SURGERY WAS PERFORMED ON THE SAME DAY TO IMPLANT A COMBINATION OF FOLLOWING PRIMA HUMERAL COMPONENTS AND DJO ALTIVATE REVERSE GLENOID: PRIMA SHORT STEM PLUS #1 (PN 1357.14.228, LOT 2524131, STERILIZATION (B)(6)) PRIMA REVERSE TRAY #SHORT (PN 1367.15.010, LOT 2518444, STERILIZATION (B)(6)) PRIMA LIMAVIT REVERSE INSERT #SHORT DIA 36MM CORR 0° (PN 1367.54.100, LOT 25AT2RE, STERILIZATION (B)(6)). DJO MODULAR BASEPLATER AND TAPER KIT (PN 508-65-001). DJO RSP GLENOID HEAD AND RETAINING SCREW 2MM LATERAL OFFSET SZ 36MM (PN 508-36-103). PATIENT DISLOCATED AND THE ABOVE MENTIONED PRIMA REVERSE TRAY AND LINER WERE REPLACED: SAME SIZE OF TRAY WAS CHOSEN BY SURGEON, WHILE THE NEW LINER WAS REPLACED WITH A RETENTIVE ONE (PN 1367.54.101). SURGERY WAS COMPLETED AS INTENDED. PATIENT IS MALE, DATE OF BIRTH (B)(6) 1957. THE EVENT OCCURRED IN THE UNITED STATES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148521 PRIMA HUMERAL SYSTEM PRIMA REVERSE INSERT # SHORT DIA.36MM 0° CORR. PHX LIMACORPORATE S.P.A 1367.54.100 25AT2RE 08033390202954

Patients

Seq Age Sex Outcome Treatment
1