FDA Adverse Event
Malfunction
Summary report: N
DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 25060915
·
Received May 1, 2026
Report
- Report Number
- 3004753838-2026-143754
- Event Type
- Malfunction
- Date Received
- May 1, 2026
- Date of Event
- April 8, 2026
- Report Date
- May 11, 2026
- Product Code
- QBJ
- UDI-DI
- 00386270002518
- PMA / PMN Number
- K213919
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
(B)(4).
Additional Manufacturer Narrative · 0
(B)(4). 3004753838-2026-143754 WAS REPORTED IN ERROR, PLEASE DISREGARD THE INITIAL REPORTING OF THIS EVENT AS THIS EVENT HAS NOW BEEN DEEMED NOT REPORTABLE.
Description of Event or Problem · 0
IT WAS REPORTED THAT ERROR ICON DISPLAY OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Description of Event or Problem · 0
SUBSEQUENT TO THE INITIAL MDR, IT WAS DETERMINED THAT A REPORT WAS SUBMITTED IN ERROR. UPON FURTHER REVIEW. IT WAS DETERMINED THAT THE PATIENT ALLEGATION DOES NOT MEET CRITERIA OF A REPORTABLE EVENT
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 165691 | DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | QBJ | 17856623 | 00386270002518 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Male |