ANGIOSCULPT EVO RX PTCA SCORING BALLOON CATHETER WITH HYDROPHILIC COATING

MDR report key: 25060244 · Received May 1, 2026

Report

Report Number
3005462046-2026-00015
Event Type
Injury
Date Received
May 1, 2026
Date of Event
April 12, 2026
Report Date
May 1, 2026
Manufacturer
PHILIPS IMAGE GUIDED THERAPY CORPORATION
Product Code
NWX
UDI-DI
00813132026899
PMA / PMN Number
P050018
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

BLOCK G2: FOREIGN- SAUDI ARABIA. BLOCK H3/H6: BASED ON THE DEVICE PHOTO RECEIVED, THE SCORING ELEMENT DETACHED FROM INTERMEDIATE BOND, BUT REMAINED INTACT TO THE DEVICE. THE ANGIOSCULPT CATHETER WAS NOT RETURNED, THUS THE CAUSE OF THE DAMAGED SCORING ELEMENT COULD NOT BE ESTABLISHED. PER THE IFU, DISSECTION IS LISTED AS A POSSIBLE ADVERSE EFFECT OF THE PROCEDURE. SUBMISSION OF THIS REPORT DOES NOT, IN ITSELF, REPRESENT A CONCLUSION BY THE MANUFACTURER AND/OR AUTHORIZED REPRESENTATIVE OR THE NATIONAL COMPETENT AUTHORITY THAT THE CONTENT OF THIS REPORT IS COMPLETE OR ACCURATE, THAT THE MEDICAL DEVICE(S) LISTED FAILED IN ANY MANNER AND/OR THAT THE MEDICAL DEVICE(S) CAUSED OR CONTRIBUTED TO AN ALLEGED DEATH OR DETERIORATION IN THE STATE OF THE HEALTH OF ANY PERSON.

Description of Event or Problem · 0

THE ANGIOSCULPT DEVICE WAS USED FOR A THERAPEUTIC CORONARY PROCEDURE IN A SEVERELY CALCIFIED PROXIMAL LAD. DURING USE, THE BALLOON WAS FOUND HANGING INSIDE THE PATIENT'S BODY, LEADING TO A DISSECTION. BASED ON THE DEVICE PHOTO RECEIVED, THE SCORING ELEMENT DETACHED FROM THE INTERMEDIATE BOND, BUT REMAINED INTACT TO THE DEVICE. THIS ADVERSE EVENT AND PRODUCT PROBLEM IS BEING SUBMITTED DUE TO DETACHED SCORING ELEMENT; RESULTED IN A DISSECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
84555 ANGIOSCULPT EVO RX PTCA SCORING BALLOON CATHETER WITH HYDROPHILIC COATING CATHETER, PTCA, CUTTING/SCORING NWX PHILIPS IMAGE GUIDED THERAPY CORPORATION 2200-3015-B 0303692519 00813132026899

Patients

Seq Age Sex Outcome Treatment
1