FDA Adverse Event
Malfunction
Summary report: N
I-STAT PT/INR CARTRIDGES
MDR report key: 2505742
·
Received March 20, 2012
Report
- Report Number
- 2245578-2012-00168
- Event Type
- Malfunction
- Date Received
- March 20, 2012
- Date of Event
- March 19, 2012
- Report Date
- March 19, 2012
- Manufacturer
- ABBOTT POINT OF CARE, INC.
- Product Code
- GJS
- PMA / PMN Number
- K020355
- Removal / Correction Number
- APOC2012-003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DETAILS WILL BE PROVIDED WITH THE ACTION THAT WILL BE CONDUCTED IN COOPERATION WITH THE U.S. FOOD AND DRUG ADMINISTRATION. (B)(4).
Description of Event or Problem · 1
A REMEDIAL ACTION HAS BEEN INITIATED TO PROVIDE RECOMMENDATIONS TO THE CUSTOMER FOR PRODUCT IN THE FIELD. ABBOTT POINT OF CARE INC. (APOC) HAS DETERMINED THAT STAT PT/INR CARTRIDGES HAVE THE POTENTIAL TO EXHIBIT INCORRECTLY ELEVATED RESULTS. INTERNAL STUDIES HAVE DEMONSTRATED THAT I-STAT PT/INR RESULTS ARE ELEVATED BY AN AVERAGE OF APPROXIMATELY 20% IN THE THERAPEUTIC RANGE OF 1.8 TO 3.5 INR. THIS ACTION IS BEING CONDUCTED IN COOPERATION WITH THE U.S. FOOD AND DRUG ADMINISTRATION AND HEALTH (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT PT/INR CARTRIDGES | PT/INR CARTRIDGES | GJS | ABBOTT POINT OF CARE, INC. | NA | ALL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |