TRUE METRIX
Report
- Report Number
- 1000113657-2026-00185
- Event Type
- Malfunction
- Date Received
- May 1, 2026
- Date of Event
- April 3, 2026
- Report Date
- May 1, 2026
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- UDI-DI
- 00021292007836
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
INTERNAL REPORT REFERENCE NUMBER: (B)(4). TEST STRIPS WERE NOT RETURNED FOR EVALUATION. METER WAS RETURNED: REPORTED DEFECT NOT REPRODUCED ON RETURNED METER. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIPS TESTED WITHIN SPECIFICATIONS MOST LIKELY UNDERLYING ROOT CAUSE: MLC-058: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 22-DEC-2020 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.
CONSUMER REPORTED COMPLAINT FOR HIGH AND ERRATIC BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER CALLED CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 214, 203, 248 AND 227 MG/DL. THE CUSTOMER STATED HER EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 104-115 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK-TO-BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. PER CUSTOMER, THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 04/29/2027 AND PER THE CUSTOMER THE OPEN VIAL DATE IS 1 MONTH. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 214 MG/DL, DATE: ON (B)(6), TIME: 2:51AM FASTING. RESULT 2: 203 MG/DL, DATE: ON (B)(6), TIME: 5:09PM UNKNOWN. RESULT 3: 248 MG/DL, DATE: ON (B)(6), TIME: 10:45AM FASTING. RESULT 4: 227 MG/DL, DATE: ON (B)(6), TIME: 10:43AM FASTING. RESULT 5: 107 MG/DL, DATE: ON (B)(6), TIME: 9:37AM FASTING
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320339 | TRUE METRIX | SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER | NBW | TRIVIDIA HEALTH INC | STRIP, TMX KGR 100CTMG/DL | ZD6235S | 00021292007836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |