DAVINCI XI
Report
- Report Number
- 2955842-2026-23505
- Event Type
- Malfunction
- Date Received
- April 30, 2026
- Date of Event
- April 9, 2026
- Report Date
- May 21, 2026
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110898
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- 003
Narratives
AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. FSE REPLACED THE REDUNDANT POWER TRAY ASSEMBLY (RPTA) DUE TO ERROR 86. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. THE COMPLAINT WAS CONFIRMED BASED ON FIELD EVALUATION. ISI HAS RECEIVED THE RPTA FOR EVALUATION, BUT EVALUATION HAS NOT BEEN COMPLETED AS OF THE DATE OF THIS REPORT. A FOLLOW-UP MDR WILL BE SUBMITTED, IF ADDITIONAL INFORMATION IS OBTAINED. PROBABLE CAUSE OF ERROR 86 POINTS TO POWER DISTRIBUTION BOARD ADC FAULT - THE MSC SAW AN ANALOG TO DIGITAL CONVERTER VOLTAGE OUT-OF-RANGE ON CHANNEL 9, 12.0V
IT WAS REPORTED THAT PRIOR TO STARTING (PRE-ANESTHESIA) A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE SYSTEM WAS PRESENTING NON-RECOVERABLE FAULT 86. PRIOR TO CALLING TECHNICAL SUPPORT ENGINEER (TSE), THE CUSTOMER HAD POWER CYCLED THE SYSTEM SEVERAL TIMES, WITH ISSUE PERSISTING. AS SYSTEM WAS ONSITE, TSE COULD REVIEW THE LOGS AND FOUND THAT IT WAS POINTING TO A VOLTAGE ISSUE INSIDE OF THE VISION SIDE CART (VSC) AND INFORMED CUSTOMER THAT IT WAS NOT RECOMMENDED TO USE THE SYSTEM IN THIS STATE, THOUGH IT WOULD BE A SURGICAL DECISION IF THEY DECIDED TO. THE CUSTOMER WOULD NOT PROCEED WITH SURGERY AS THERE WAS NO PATIENT IN THE THEATRE. THE PROCEDURE WAS COMPLETED AS PLANNED WITH NO REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 207273 | DAVINCI XI | VISION SIDE SYSTEM | NAY | INTUITIVE SURGICAL, INC | 381121-46 | N/A | 00886874110898 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |