Report

Report Number
2249529-2026-00001
Event Type
Malfunction
Date Received
April 28, 2026
Report Date
April 27, 2026
Manufacturer
GSOURCE, LLC.
Product Code
HTW
UDI-DI
00840314246721
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING DETAILS THE INVESTIGATION OF A BREAKAGE OF TWIST DRILL 5 INCH STRAIGHT SHANK DOUBLE FLUTED DIAMETER 2.0MM 23MM ROUND END. BASED ON THE RECEIVED INFORMATION, THE DRILL BIT SIZE WAS USED IN A PROCEDURE INVOLVING THE FEMUR. THE LIKELY ROOT CAUSE OF THE BREAKAGE IS THE USAGE OF A DRILL BIT DIAMETER (Ø 2.0MM) THAT IS TOO THIN FOR USAGE IN THE FEMUR DUE TO ITS HIGH BONE DENSITY. USING THIS THIN DIAMETER ON THE FEMUR ALSO MAKES THE DRILL MORE SUSCEPTIBLE TO BREAKAGE IF EXCESSIVE FORCE IS APPLIED OR FORCE APPLIED AT AN INCORRECT ANGLE. IN 2025, GSOURCE QUALITY AND ENGINEERING HAVE ALREADY IMPROVED THE DRILL BIT DESIGN TO CORRECT/PREVENT BREAKAGES AND HAVE MADE A GENERAL INSTRUCTIONS FOR USE ASSISTING HEALTHCARE PROFESSIONALS WITH THE SELECTION AND PROPER USAGE OF DRILL BITS FOR MEDICAL PROCEDURES.

Description of Event or Problem · 0

A DRILL BIT BROKE OFF INTO THE PATIENT'S FEMUR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
231283 GSOURCE TWIST DRILL 5" STR SHANK DOUBLE FLUTED Ø 2.0MM 23MM ROUND END HTW GSOURCE, LLC. GS 86.8420 5232541 00840314246721

Patients

Seq Age Sex Outcome Treatment
1