O-ARM O2 IMAGING SYSTEM
Report
- Report Number
- 3004785967-2026-00214
- Event Type
- Malfunction
- Date Received
- April 28, 2026
- Date of Event
- January 26, 2026
- Report Date
- April 28, 2026
- Manufacturer
- MEDTRONIC NAVIGATION, INC. (BOXBOROUGH)
- Product Code
- OWB
- UDI-DI
- 00763000616342
- PMA / PMN Number
- K200074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: BI71000610, SERIAL/LOT #: -, UBD: , UDI#: H3: A MANUFACTURER REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. IN SUMMARY, THE CABLE FOR THE Z DRIVE WAS DAMAGED BY THE G ANTRY AND WAS SHORTED. THE Z DRIVE MOTOR CABLE WAS REPLACED. A SYSTEM CHECKOUT WAS PERFORMED, AND THE SYSTEM PERFORMED AS INTENDED. CODES FDM B01, FDR C0208, FDC D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
MEDTRONIC RECEIVED INFORMATION REGARDING AN IMAGING SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE SYSTEM DID NOT INITIALIZE MOTION. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 173041 | O-ARM O2 IMAGING SYSTEM | INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM | OWB | MEDTRONIC NAVIGATION, INC. (BOXBOROUGH) | BI70002000 | 00763000616342 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |