FDA Adverse Event Malfunction Summary report: N

RITTER

MDR report key: 2501683 · Received February 28, 2012

Report

Report Number
1523530-2012-00008
Event Type
Malfunction
Date Received
February 28, 2012
Date of Event
February 2, 2012
Report Date
February 27, 2012
Manufacturer
MIDMARK CORP.
Product Code
FQO
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE TABLE WAS EXAMINED AT THE FACILITY BY A MIDMARK REPRESENTATIVE. EXAMINATION OF THE TABLE INDICATES THE DRAWER FRONT WAS HANGING PAST ITS INTENDED POSITION AND THE TABLE WAS LOWERED ONTO IT. WHEN THE TABLE WAS LOWERED, THE DRAWER FRONT WAS PINCHED BETWEEN THE FLOOR AND THE DRAWER UNTIL THE DRAWER FRONT MOVED. WHEN THE DRAWER FRONT MOVED, THE TABLE DROPPED A FEW INCHES VERY QUICKLY. THE DRAWER FRONT IS DESIGNED TO OPEN AS A SHELF. IT IS ALSO DESIGNED SO THAT IT CAN BREAK AWAY IF A PATIENT WERE TO EXIT THE TABLE WITH THE DRAWER IN THE DOWN POSITION. THE DRAWER FRONT MUST THEN BE REPOSITIONED BY DEPRESSING THE TABS ON ITS FRONT. THIS PROCESS IS DESCRIBED IN THE PRODUCTS USER'S GUIDE. FAILURE TO CORRECTLY REPOSITION THE DRAWER FRONT CAN RESULT IN THE TYPE OF DAMAGE SEEN WITH THIS TABLE.

Description of Event or Problem · 1

UNIT WAS LOWERED ONTO THE DRAWER FRONT. THE BACK CONTINUED TO LOWER UNTIL IT REACHED A 3 INCH DECLINE WHEN THE DRAWER FRONT GAVE WAY CAUSING THE FRONT OF THE TABLE TO DROP. NO INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RITTER EXAMINATION TABLE FQO MIDMARK CORP. 222-015

Patients

Seq Age Sex Outcome Treatment
1