FDA Adverse Event
Malfunction
Summary report: N
ALINITY I FREE T4 REAGENT KIT
MDR report key: 25016385
·
Received April 28, 2026
Report
- Report Number
- 3005094123-2026-00196
- Event Type
- Malfunction
- Date Received
- April 28, 2026
- Date of Event
- April 19, 2026
- Report Date
- April 28, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- CEC
- UDI-DI
- 00380740173715
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07P70 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 07P70, WITH 510K NUMBER K173122. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION.
Description of Event or Problem · 0
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I FREE T4 FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE: 0.70-1.48 NG/DL): INITIAL FREE T4 RESULT= 1.78 NG/DL; REPEAT RESULT= 1.06 NG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 626458 | ALINITY I FREE T4 REAGENT KIT | RADIOIMMUNOASSAY, FREE THYROXINE | CEC | ABBOTT IRELAND DIAGNOSTICS DIVISION | 82597UD01 | 00380740173715 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |