AMISTEM-P COLLARED
Report
- Report Number
- 3005180920-2026-00351
- Event Type
- Injury
- Date Received
- April 28, 2026
- Date of Event
- March 31, 2026
- Report Date
- April 28, 2026
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630040720328
- PMA / PMN Number
- K173794
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
BATCH REVIEW PERFORMED ON 07 APRIL 2026 STEM: AMISTEM-P COLLARED 01.18.443 AMISTEM-P COLLARED LAT. SIZE 3 (K173794) LOT 2428259: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 23-JAN-2025. EXPIRATION DATE: 07-JAN-2030. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. CLINICAL EVALUATION LESS THAN ONE YEAR AFTER PRIMARY CEMENTLESS THA THE STEM GETS LOOSE, IT WAS PROBABLY NEVER INTEGRATED BY THE HOST BONE. NO TRAUMA REPORTED. IT SHOULD PROBABLY BE CLASSIFIED AS A CASE OF EARLY IDIOPATHIC LOOSENING, AND NO CLEAR CAUSE CAN BE DETERMINED FROM THE EVIDENCE SUPPLIED. THE REPORT MENTIONS A COLLARED AMISTEM, BUT THE IMAGE SUPPLIER REPRESENTS A COLLARLESS STEM. ROOT CAUSE: ASEPTIC LOOSENING IS POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT AFTER PRIMARY HIP ARTHROPLASTIES AND CAUSES ARE OFTEN UNKNOWN. THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AND THE INVESTIGATION DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED ISSUE OR SYSTEMIC DEFICIENCY.
AT ABOUT 9 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN PRESENTING PAIN DUE TO A LOOSE STEM AND THE CAUSE IS UNKNOWN. THERE WAS NO INDICATION OF TRAUMA. THE SURGEON MADE NO NOTE OF BONE QUALITY. THE SURGEON REVISED THE MEDACTA STEM AND HEAD TO A COMPETITOR STEM AND HEAD, AND REVISED THE MEDACTA D 40 FLAT LINER TO A D 28 DOUBLE MOBILITY LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 137993 | AMISTEM-P COLLARED | AMISTEM-P COLLARED LAT. SIZE3 | LZO | MEDACTA INTERNATIONAL SA | 01.18.443 | 2428259 | 07630040720328 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Male | Required Intervention |