FDA Adverse Event Injury Summary report: N

AMISTEM-P COLLARED

MDR report key: 25012263 · Received April 28, 2026

Report

Report Number
3005180920-2026-00351
Event Type
Injury
Date Received
April 28, 2026
Date of Event
March 31, 2026
Report Date
April 28, 2026
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630040720328
PMA / PMN Number
K173794
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 07 APRIL 2026 STEM: AMISTEM-P COLLARED 01.18.443 AMISTEM-P COLLARED LAT. SIZE 3 (K173794) LOT 2428259: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 23-JAN-2025. EXPIRATION DATE: 07-JAN-2030. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. CLINICAL EVALUATION LESS THAN ONE YEAR AFTER PRIMARY CEMENTLESS THA THE STEM GETS LOOSE, IT WAS PROBABLY NEVER INTEGRATED BY THE HOST BONE. NO TRAUMA REPORTED. IT SHOULD PROBABLY BE CLASSIFIED AS A CASE OF EARLY IDIOPATHIC LOOSENING, AND NO CLEAR CAUSE CAN BE DETERMINED FROM THE EVIDENCE SUPPLIED. THE REPORT MENTIONS A COLLARED AMISTEM, BUT THE IMAGE SUPPLIER REPRESENTS A COLLARLESS STEM. ROOT CAUSE: ASEPTIC LOOSENING IS POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT AFTER PRIMARY HIP ARTHROPLASTIES AND CAUSES ARE OFTEN UNKNOWN. THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AND THE INVESTIGATION DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED ISSUE OR SYSTEMIC DEFICIENCY.

Description of Event or Problem · 0

AT ABOUT 9 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN PRESENTING PAIN DUE TO A LOOSE STEM AND THE CAUSE IS UNKNOWN. THERE WAS NO INDICATION OF TRAUMA. THE SURGEON MADE NO NOTE OF BONE QUALITY. THE SURGEON REVISED THE MEDACTA STEM AND HEAD TO A COMPETITOR STEM AND HEAD, AND REVISED THE MEDACTA D 40 FLAT LINER TO A D 28 DOUBLE MOBILITY LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
137993 AMISTEM-P COLLARED AMISTEM-P COLLARED LAT. SIZE3 LZO MEDACTA INTERNATIONAL SA 01.18.443 2428259 07630040720328

Patients

Seq Age Sex Outcome Treatment
1 58 YR Male Required Intervention