FDA Adverse Event Injury Summary report: N

T3® PRO NON-PLATFORM SWITCHED TAPERED IMPLANT 5MM (D) X 13MM (L)

MDR report key: 25008267 · Received April 27, 2026

Report

Report Number
0001038806-2026-02313
Event Type
Injury
Date Received
April 27, 2026
Date of Event
April 10, 2026
Report Date
April 27, 2026
Manufacturer
ZIMVIE US CORP LLC
Product Code
DZE
UDI-DI
00844868046981
PMA / PMN Number
K213672
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4) A4: WEIGHT UNKNOWN / NOT PROVIDED. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NON-EXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS SEVERE BONE LOSS AROUND THE IMPLANT AT TOOTH SITE #29. THE IMPLANT WAS REMOVED. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
493994 T3® PRO NON-PLATFORM SWITCHED TAPERED IMPLANT 5MM (D) X 13MM (L) DENTAL IMPLANT DZE ZIMVIE US CORP LLC 2120014818 00844868046981

Patients

Seq Age Sex Outcome Treatment
1 70 YR Female Required Intervention