FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2500699 · Received March 22, 2012

Report

Report Number
2531779-2012-02364
Event Type
Malfunction
Date Received
March 22, 2012
Report Date
February 23, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE RETAIN CARTRIDGE FROM SAME LOT # B201724 WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED WITH NO DAMAGE OR DEFECTS NOTED. A LEAK TEST AND FILL TEST WERE PERFORMED WITH NO ISSUES FOUND. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Description of Event or Problem · 1

THE PATIENT CLAIMED THE ANIMAS CARTRIDGE HAS SMALL AIR BUBBLE ISSUES. REPORTEDLY, THE PATIENT NOTICED SMALL AIR BUBBLE WHERE THE TUBING AND CARTRIDGE CONNECT. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO THE SMALL AIR BUBBLE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN PUMP CARTRIDGE LZG ANIMAS CORPORATION ANM IR1200/1250/2020/OTP CART

Patients

Seq Age Sex Outcome Treatment
1 67 YR