FDA Adverse Event Malfunction Summary report: N

BERKETTE BEACH CHAIR LIMB POSITIONER

MDR report key: 24981539 · Received April 24, 2026

Report

Report Number
3002750084-2026-00041
Event Type
Malfunction
Date Received
April 24, 2026
Date of Event
March 27, 2026
Report Date
April 24, 2026
Manufacturer
ASPEN SURGICAL PRODUCTS, INC.
Product Code
CCX
UDI-DI
00840113230662
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED FOR EVALUATION, BUT PICTURES WERE PROVIDED SO THE COMPLAINT COULD BE CONFIRMED. THE INVESTIGATION CONCLUDED THAT THE BREAKAGE OF THE BRACKET HANDLE WAS DUE TO A DEVICATION ATTRIBUTABLE TO THE EXTERNAL SUPPLIER THAT PROVIDES THIS COMPONENT. THE SUPPLIER WAS NOTIFIED, AND CORRECTIVE ACTIONS ARE BEING FOLLOWED UP. THE ROOT CAUSE IS A MANUFACTURING ERROR. THIS SHOULD BE CONSIDERED THE FINAL REPORT. IF ANY ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 0

COMPLAINANT ALLEGES "THIS MORNING WE HAD THE HANDLE ON THE BERKETTE BREAK OFF WHILE GETTING THE PATIENT POSITIONED. THIS HAPPENED 3 TIMES TO 3 DIFFERENT PRODUCTS ALL WITH THE SAME LOT# OF 110525. HOPEFULLY YOU CAN TELL FROM THE PICTURES BELOW, BUT THE HANDLE SNAPPED WHERE IT IS COVERED BY THE GRIP."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
246060 BERKETTE BEACH CHAIR LIMB POSITIONER PATIENT POSITIONING CCX ASPEN SURGICAL PRODUCTS, INC. 711200 110525 00840113230662

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown