BERKETTE BEACH CHAIR LIMB POSITIONER
Report
- Report Number
- 3002750084-2026-00041
- Event Type
- Malfunction
- Date Received
- April 24, 2026
- Date of Event
- March 27, 2026
- Report Date
- April 24, 2026
- Manufacturer
- ASPEN SURGICAL PRODUCTS, INC.
- Product Code
- CCX
- UDI-DI
- 00840113230662
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS NOT RETURNED FOR EVALUATION, BUT PICTURES WERE PROVIDED SO THE COMPLAINT COULD BE CONFIRMED. THE INVESTIGATION CONCLUDED THAT THE BREAKAGE OF THE BRACKET HANDLE WAS DUE TO A DEVICATION ATTRIBUTABLE TO THE EXTERNAL SUPPLIER THAT PROVIDES THIS COMPONENT. THE SUPPLIER WAS NOTIFIED, AND CORRECTIVE ACTIONS ARE BEING FOLLOWED UP. THE ROOT CAUSE IS A MANUFACTURING ERROR. THIS SHOULD BE CONSIDERED THE FINAL REPORT. IF ANY ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.
COMPLAINANT ALLEGES "THIS MORNING WE HAD THE HANDLE ON THE BERKETTE BREAK OFF WHILE GETTING THE PATIENT POSITIONED. THIS HAPPENED 3 TIMES TO 3 DIFFERENT PRODUCTS ALL WITH THE SAME LOT# OF 110525. HOPEFULLY YOU CAN TELL FROM THE PICTURES BELOW, BUT THE HANDLE SNAPPED WHERE IT IS COVERED BY THE GRIP."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 246060 | BERKETTE BEACH CHAIR LIMB POSITIONER | PATIENT POSITIONING | CCX | ASPEN SURGICAL PRODUCTS, INC. | 711200 | 110525 | 00840113230662 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |