FDA Adverse Event Malfunction Summary report: N

INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT

MDR report key: 2497735 · Received March 20, 2012

Report

Report Number
9611451-2012-00160
Event Type
Malfunction
Date Received
March 20, 2012
Date of Event
February 29, 2012
Report Date
February 29, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT265 INFANT EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. METHOD: THE COMPLAINT RT265 INFANT BREATHING CIRCUIT HAS BEEN RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) AND PERFORMANCE TESTED. RESISTANCE AND SURFACE TEMPERATURE MEASUREMENTS WERE PERFORMED ON THE INSPIRATORY AND EXPIRATORY LIMBS OF THE COMPLAINT DEVICE. RESULT: THE RESISTANCE AND SURFACE TEMPERATURE MEASUREMENTS OF THE INSPIRATORY AND EXPIRATORY LIMBS WERE WITHIN SPECIFICATIONS. THE HOSPITAL ALSO REPORTED THAT THE BREATHING TUBE WAS LYING DIRECTLY ON THE PATIENT'S LEG, WITHOUT BEING COVERED. A LOT CHECK WAS NOT PERFORMED AS NO LOT NUMBER WAS PROVIDED. CONCLUSION: WE WERE UNABLE TO CONFIRM THE REPORTED FAULT OF THE BREATHING CIRCUIT BECOMING VERY HOT. NO FAULT WAS FOUND ON THE RETURNED COMPLAINT DEVICE. THE REPORTED TEMPORARY REDDENING OF THE PATIENT'S THIGH WAS MOST LIKELY CAUSED BY PROLONGED CONTACT OF THE BREATHING TUBE WITH THE PATIENT'S LEG. THE USER INSTRUCTIONS PROVIDED WITH THE RT265 INFANT BREATHING CIRCUIT STATES: - "AVOID PROLONGED CONTACT WITH PATIENT'S SKIN." - "DO NOT COVER THE CIRCUIT WITH MATERIALS SUCH AS BLANKETS, TOWELS OR BED LINEN."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT AN RT265 INFANT BREATHING CIRCUIT KIT BECAME VERY HOT AND CAUSED TEMPORARY REDDENING ON THE PATIENT'S THIGH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT265

Patients

Seq Age Sex Outcome Treatment
1