FDA Adverse Event Malfunction Summary report: N

INSYTE AUTOGUARD

MDR report key: 24970100 · Received April 23, 2026

Report

Report Number
1710034-2026-00465
Event Type
Malfunction
Date Received
April 23, 2026
Date of Event
April 7, 2026
Report Date
April 28, 2026
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825233
PMA / PMN Number
SEE H.11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. G.4. K201075; K251654. H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

THE COMPLAINT OF NEEDLE RETRACTION ISSUES COULD NOT BE CONFIRMED FROM THE REPRESENTATIVE 22G INSYTE AUTOGUARD UNITS THAT WERE RECEIVED IN SEALED PACKAGING FROM LOT #5252921. A FUNCTIONAL TEST REVEALED THAT THE NEEDLES FULLY RETRACTED WITHOUT RESISTANCE OR DELAY WHEN THE SAFETY MECHANISM WAS ACTIVATED. A REVIEW OF THE INSPECTION RECORDS AND QUALITY AND MANUFACTURING CONTROLS FOR THE IMPLICATED LOT INDICATED NO ISSUES WITH THE MANUFACTURING PROCESS. ALTHOUGH THE REPORTED ISSUE COULD NOT BE CONFIRMED, THE COMPLAINT HAS BEEN DOCUMENTED AND WILL CONTINUE TO BE MONITORED AS PART OF ONGOING EFFORTS TO IDENTIFY POTENTIAL MANUFACTURING RELATED ISSUES. A REVIEW OF OUR RISK MANAGEMENT DOCUMENTATION WAS PERFORMED AND INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY. COMPLAINTS RECEIVED ABOUT THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE DATA COLLECTED FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE NEEDLES ARE NOT RETRACTING. (B)(4): NEEDLES ARE NOT RETRACTING, SENDING BACK WHAT THEY HAVE.

Description of Event or Problem · 0

NO NEW INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
615088 INSYTE AUTOGUARD PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 5252921 00382903825233

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown