FDA Adverse Event Malfunction Summary report: N

CARESTATION 750

MDR report key: 24967913 · Received April 23, 2026

Report

Report Number
9710602-2026-01128
Event Type
Malfunction
Date Received
April 23, 2026
Date of Event
March 27, 2026
Report Date
April 23, 2026
Manufacturer
GE MEDICAL SYSTEMS (CHINA) CO., LTD
Product Code
BSZ
UDI-DI
00840682145596
PMA / PMN Number
K213867
Removal / Correction Number
SEE H11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RO
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

GE HEALTHCARE (GEHC) REPORTED A FIELD MODIFICATION FOR AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE PER 21 CFR 806 ON 02-APR-2025. THE FDA RECALL NUMBERS ARE: Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025. CUSTOMERS WERE SENT A LETTER EXPLAINING THE ISSUE AND PROVIDING INSTRUCTIONS FOR INSPECTING FOR EFFECTIVE VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. GEHC WILL CORRECT ALL DEVICES THAT FAIL THE VENTILATION SCREENING TEST AT NO COST. BLOCK A: NO PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS MADISON - 3030 OHMEDA DR, USA MADISON, WI 53718.

Description of Event or Problem · 0

AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
283537 CARESTATION 750 ANESTHESIA GAS MACHINE BSZ GE MEDICAL SYSTEMS (CHINA) CO., LTD A1 NA 00840682145596

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown