IMPELLA
Report
- Report Number
- 1220648-2026-07073
- Event Type
- Injury
- Date Received
- April 23, 2026
- Date of Event
- April 18, 2026
- Report Date
- April 23, 2026
- Manufacturer
- ABIOMED, INC. - 1220648
- Product Code
- OZD
- UDI-DI
- 00813502012279
- PMA / PMN Number
- P140003
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ABIOMED INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ABIOMED INC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
H6: UPDATED CODES BASED ON THE RESULTS OF THE INVESTIGATION. H11: UPDATED BASED ON THE RESULTS OF THE INVESTIGATION AND ADDED CLINICAL REVIEW. CORRECTED INFORMATION HAS BEEN PROVIDED IN D3( MANUFACTURER FAX); D4(SERIAL); E4. THE CAUSE OF THE INJURY WAS NOT DETERMINED DUE TO INSUFFICIENT CLINICAL DETAILS.
AN IMPELLA CP DEVICE WAS INSERTED INTO THE RIGHT FEMORAL ARTERY IN A 60-YEAR-OLD MALE PATIENT WITH A PAST MEDICAL HISTORY OF CORONARY ARTERY DISEASE (CAD), PRESENTING IN ACUTE MYOCARDIAL INFARCTION (AMI) AND CARDIOGENIC SHOCK (CGS), IN AN OUT-OF-HOSPITAL CARDIAC ARREST REQUIRING CARDIOPULMONARY RESUSCITATION (CPR), IN SCAI STAGE E SHOCK, ON MULTIPLE INOTROPES AND VASOPRESSORS, AND ON A VENTILATOR FOR RESPIRATORY SUPPORT, PRIOR TO INITIATION OF SUPPORT. THE IMPELLA WAS INSERTED FOR VENTRICULAR SUPPORT FOR A PROTECTED PERCUTANEOUS CORONARY INTERVENTION (PCI). WHILE THE PATIENT WAS IN THE INTENSIVE CARE UNIT (ICU), THERE WAS HEMATURIA. THE P-LEVEL WAS LOWERED FROM P-9 AT P-4, WHICH RESOLVED THE ISSUE. THE URINE WAS CLEAR IN THE MORNING. AFTER ONE DAY ON SUPPORT, THE DEVICE WAS SUCCESSFULLY WEANED. THE POST-PROCEDURE OUTCOME IS SURVIVED AT EXPLANT. THE IMPELLA FUNCTIONED AT P-4 AT 2.3L/MIN AS INTENDED. BLEEDING (HEMATURIA) IS A KNOWN RISK DUE TO THE IMPELLA'S ANTICOAGULATION AND PURGE REQUIREMENTS. THE PUMP IS NOT AVAILABLE FOR RETURN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 180845 | IMPELLA | TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP | OZD | ABIOMED, INC. - 1220648 | 2027854451 | 00813502012279 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |