FDA Adverse Event
Other
Summary report: N
KARL STORZ
MDR report key: 2496457
·
Received March 15, 2012
Report
- Report Number
- 9610617-2012-00009
- Event Type
- Other
- Date Received
- March 15, 2012
- Date of Event
- November 21, 2011
- Report Date
- March 13, 2012
- Manufacturer
- KARL STORZ GMBH & CO. KG
- Product Code
- GCJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
HOSP REPORTED THAT DOCTOR TORQUED SCOPE, AND THE LENS CRACKED. SCOPE EVAL CONFIRMED THAT SCOPE HAD BROKEN ROD LENSES.
Description of Event or Problem · 1
ALLEGEDLY, DURING A TUR OF BLADDER TUMOR, DOCTOR INSERTED SCOPE IN URETHRA AND TORQUED IT WHILE POSITIONING THE SCOPE, AND THE LENS CRACKED. HE HAD NO FURTHER VISUALIZATION AND HAD TO ABORT THE PROCEDURE. PROCEDURE WAS RESCHEDULED FOR A LATER DATE AND WAS COMPLETED WITH NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KARL STORZ | HOPKINS TELESCOPE | GCJ | KARL STORZ GMBH & CO. KG | 27325BA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Other |