FDA Adverse Event
Malfunction
Summary report: N
ABC SSP UNITRAY KIT WITH TAQ
MDR report key: 2495453
·
Received March 8, 2012
Report
- Report Number
- 2244574-2012-00002
- Event Type
- Malfunction
- Date Received
- March 8, 2012
- Date of Event
- December 12, 2011
- Report Date
- March 7, 2012
- Manufacturer
- LIFE TECHNOLOGIES CORPORATION
- Product Code
- IYO
- PMA / PMN Number
- BK020068
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
ONE COMPLAINT HAS BEEN RECEIVED. ABC UNITRAY SSP KITS HAVE A FALSE NEGATIVE IN LANE 13. THE CUSTOMER REPORTED THAT, IN COMPARISON WITH A COMPETITOR'S SSP PRODUCT, THE ABC SSP UNITRAY WAS NEGATIVE IN LANE 13 FOR A TYPING OF A 32:04. LANE 13 WAS LISTED ON THE KIT WORKSHEET/INTERPRETATION DOCUMENTATION (WHICH IS INCLUDED WITH EACH KIT) AS THE FOLLOWING: A 32:04^, HOWEVER THE DOCUMENTATION ALSO INDICATED THAT "ALLELES WRITTEN WITH THE "^" SYMBOL FOLLOWING THEM HAVE UNK SEQUENCES AT THE SENSE AND / OR ANTISENSE PRIMER ANNEALING SITE. THEIR REACTION PATTERNS ARE PREDICTED BASED ON SEQUENCE HOMOLOGY WITH EACH MAJOR GROUP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ABC SSP UNITRAY KIT WITH TAQ | MZI TEST, QUALITATIVE FOR HLA, NON-DIAGN | IYO | LIFE TECHNOLOGIES CORPORATION | 7800110 | 037 979628 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |