FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SERUM BLOOD COLLECTION TUBES

MDR report key: 24947993 · Received April 21, 2026

Report

Report Number
1024879-2026-00599
Event Type
Malfunction
Date Received
April 21, 2026
Date of Event
March 25, 2026
Report Date
March 30, 2026
Manufacturer
BECTON DICKINSON & CO (FRANKLIN LAKES)
Product Code
JKA
UDI-DI
30382903678144
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K960250. INVESTIGATION SUMMARY: BD RECEIVED 2 PHOTOS FOR INVESTIGATION, WHICH SHOW TWO SAMPLES THAT ARE ONLY HALFWAY FULL. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE: DRAW VOLUME. NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM BLOOD COLLECTION TUBES, TWO (2) TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1008467 BD VACUTAINER® SERUM BLOOD COLLECTION TUBES TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION JKA BECTON DICKINSON & CO (FRANKLIN LAKES) 5105267 30382903678144

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown