LINK MP TRIAL NECK SEGMENT
Report
- Report Number
- 9610548-2012-00008
- Event Type
- Other
- Date Received
- March 8, 2012
- Date of Event
- December 19, 2011
- Report Date
- March 8, 2012
- Manufacturer
- WALDEMAR LINK GMBH & CO. KG
- Product Code
- JDI
- PMA / PMN Number
- K955296
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- NOT APPLICABLE
Narratives
WE STARTED A VOLUNTARY PRODUCT RECALL IN THE COUNTRIES WHICH WE DELIVERED WITH THE AFFECTED PRODUCTS ((B)(4)). THE INFO OF THE RECALL IS ENCLOSED. THIS EVENT OCCURRED OUTSIDE OF THE U.S AND INVOLVED A PRODUCT THAT WAS MFG OUTSIDE OF THE U.S. HOWEVER, BECAUSE THE LINK MP RECONSTRUCTION SYSTEM ALSO IS MARKETED IN THE U.S LINK IS SUBMITTING THIS MDR TO ENSURE FULL COMPLIANCE WITH 21 CFR PART 803.
THE CUSTOMER OPENED A 131-394/35 TRIAL NECK SEGMENT MARKED AS CCD 135 DEGREES, BUT IN FACT THE ONE IN THE BOX IS A CCD 126 DEGREES. (B)(4) 2011: OUR DISTRIBUTOR (B)(4), INFORMED US THAT A CUSTOMER OPENED A 131-394/35 TRIAL NECK SEGMENT MARKED AS CCD 135 DEGREES, BUT IN FACT THE ONE IN THE BOX IS A CCD 126 DEGREES. AT THIS TIME NO FURTHER INVESTIGATION POSSIBLE BECAUSE WE DIDN'T RECEIVE THE LOT NUMBER FROM THE USER. FEB 6TH, 2012: COMPLAINT SAMPLE RECEIVED AND JUSTIFIED. ACCORDING TO OUR DOCUMENTATION TWO BATCHES WERE MIXED UP DURING THE PRODUCTION PROCESS AND FINALLY INCORRECTLY LABELED. REF 131-394/35 LOT: A911205 10 PIECES: LABELLED CCD ANGLE 135 DEGREES, PHYSICAL CCD ANGLE 126 DEGREES. REF: 131-394/26 LOT: A911179 10 PIECES: LABELLED CCD ANGLE 126 DEGREES, PHYSICAL CCD ANGLE 135 DEGREES. DUE TO THE FACT THAT NO ITEM OF THE COMPLEMENT BATCH ((B)(4)) WAS ON STOCK WE NEEDED TO CHECK IT AT CUSTOMER SITE, WHICH WAS CONFIRMED ON 24TH FEB, 2012 BY OUR DISTRIBUTOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LINK MP TRIAL NECK SEGMENT | MP TRIAL NECK SEGMENT | JDI | WALDEMAR LINK GMBH & CO. KG | 131-394/35 | A911205 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |