FDA Adverse Event Injury Summary report: N

1.25MM KIRSCHNER WIRE, TROCAR POINTS ON BOTH ENDS

MDR report key: 2491439 · Received March 7, 2012

Report

Report Number
3003506883-2012-00044
Event Type
Injury
Date Received
March 7, 2012
Date of Event
December 2, 2011
Report Date
December 2, 2011
Manufacturer
SYNTHES ELMIRA
Product Code
HXI
PMA / PMN Number
PREAMEND
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE RAW MATERIAL DEVICE HISTORY RECORD REVEALED THIS LOT MET ALL SPECIFICATIONS WITH NO ANOMALIES NOTED. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO COMPLAINT RELATED ANOMALIES. THIS ORDER MET ALL DIMENSIONAL AND VISUAL CRITERIA AT THE OF RELEASE WITH NO ISSUES DOCUMENTED DURING THE MANUFACTURE THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE WIRES CONFORM TO ALL DIMENSIONAL SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY REVEALED NO COMPLAINT RELATED ANOMALIES. THIS LOT MET ALL DIMENSIONAL AND VISUAL CRITERIA AT THE TIME OF MANUFACTURE WITH NO ISSUES DOCUMENTED DURING THE MANUFACTURE THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. REVIEW OF THE PACKAGING DEVICE HISTORY SHOWS THAT THESE WIRES WERE PACKAGED IN A 10 PIECE PACKAGE WITH TIP PROTECTORS AT BOTH ENDS. THIS LOT WAS PACKAGED WITH NO ISSUES DOCUMENTED THAT WOULD LEAD TO THE COMPLAINT CONDITION.

Description of Event or Problem · 1

ACCOUNT REPORTED TO CONSULTANT: DURING INVENTORY CONTROL, AFTER COUNTING AND WHILE PUTTING THE K-WIRE PACKAGE BACK, ONE END OF THE K-WIRE PUNCTURED THROUGH THE PALM OF THE AND THE OTHER END PUNCTURED THROUGH THE FINGER OF HOSPITAL PERSONNEL. HOSPITAL PERSONNEL HAS BEEN BANDAGED UP AND THE BLEEDING HAS STOPPED. THE (3) K-WIRES PUNCTURED THE HOSPITAL PERSONNEL. THE (3) K-WIRES PUNCTURED THE RIGHT HAND AND (2) PUNCTURED THE LEFT HAND. THIS IS 2 OF 3 REPORTS FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 1.25MM KIRSCHNER WIRE, TROCAR POINTS ON BOTH ENDS 1.25MM KIRSCHNER WIRE, TROCAR POINTS ON BOTH ENDS HXI SYNTHES ELMIRA 6696566

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention