THERMOCOOL SMARTTOUCH SF
Report
- Report Number
- 2029046-2026-01244
- Event Type
- Malfunction
- Date Received
- April 16, 2026
- Date of Event
- February 8, 2026
- Report Date
- April 16, 2026
- Manufacturer
- BIOSENSE WEBSTER INC
- Product Code
- LPB
- UDI-DI
- 10846835009804
- PMA / PMN Number
- P030031/S078
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
THE ¿SUSPECTED MEDICAL DEVICE¿ REPORTED IN SECTION D OF THIS REPORT IS NOT MARKETED IN USA OR APPROVED BY THE FDA. HOWEVER, IT IS BEING REPORTED AS BIOSENSE WEBSTER CONSIDERS THIS AS A SIMILAR DEVICE TO THERMOCOOL SMARTTOUCH SF CATHETER APPROVED UNDER P030031/S078. THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE EVALUATION DETAILS: THE DEVICE WAS RETURNED TO JOHNSON & JOHNSON MEDTECH (J&J MEDTECH) FOR EVALUATION. FOLLOWING J&J MEDTECH PROCEDURES, A VISUAL INSPECTION AND TEMPERATURE AND IMPEDANCE TEST OF THE RETURNED DEVICE WERE PERFORMED. VISUAL ANALYSIS REVEALED REDDISH BROWN MATERIAL INSIDE THE PEBAX AND A SEPARATION BETWEEN PEBAX AND ELECTRODE SECTION. A TEMPERATURE AND IMPEDANCE TEST WAS PERFORMED AND NO TEMPERATURE VALUES WERE DISPLAYED. THE THERMOCOUPLE CONTINUITY WAS MEASURED, AND THE RESULTS WERE FOUND WITHIN SPECIFICATIONS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE 31606488L NUMBER, AND NO INTERNAL ACTIONS RELATED TO THE REPORTED COMPLAINT CONDITION WERE IDENTIFIED. THE REDDISH MATERIAL INSIDE THE PEBAX COULD BE RELATED TO THE TEMPERATURE SENSOR ISSUE REPORTED BY THE CUSTOMER; THEREFORE, THE CUSTOMER COMPLAINT WAS CONFIRMED. THE ROOT CAUSE OF THE PEBAX DAMAGE IS TRACED TO THE MANUFACTURING PROCESS. THE INSTRUCTIONS FOR USE (IFU) CONTAIN THE FOLLOWING RECOMMENDATIONS: WHEN CLEANING THE TIP ELECTRODE, BE CAREFUL NOT TO TWIST THE TIP ELECTRODE WITH RESPECT TO THE CATHETER SHAFT; TWISTING MAY DAMAGE THE TIP ELECTRODE BOND AND LOOSEN THE TIP ELECTRODE, OR MAY DAMAGE THE CONTACT FORCE SENSOR. A SIGNIFICANT CHANGE IN THE BASELINE READING AFTER CLEANING MIGHT INDICATE A DAMAGED CONTACT FORCE SENSOR. IN ADDITION, IN ORDER TO PREVENT DAMAGE TO THE CATHETER TIP, VERIFY COMPATIBILITY BETWEEN THE SHEATH AND CATHETER, ADVANCE THE CATHETER THROUGH THE SHEATH PRIOR TO INSERTION. ANY SHEATH < 8.5 F IS CONTRAINDICATED. THIS PRODUCT ISSUE WILL BE ADDRESSED THROUGH THE QUALITY SYSTEM. AN INTERNAL ACTION WAS CREATED FOR FURTHER INVESTIGATION OF THE FORCE SENSOR SLEEVE INSOLATION TO PREVENT FLUIDS TO GET INSIDE INTO THE DEVICE. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION - PAROXYSMAL ABLATION PROCEDURE WITH A THERMOCOOL SMARTTOUCH SF CATHETER AND POST PROCEDURE THE BWI PRODUCT ANALYSIS LAB SEPARATION BETWEEN THE PEBAX AND ELECTRODE SECTION. DURING THE PROCEDURE ITSELF, THERE WAS NO TEMPERATURE DISPLAYED ON THE CARTO OR GENERATOR. A SECOND DEVICE WAS USED TO COMPLETE THE OPERATION. THERE WAS NO ADVERSE EVENT REPORTED ON PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 961917 | THERMOCOOL SMARTTOUCH SF | CARDIAC ABLATION PERCUTANEOUS CATHETER | LPB | BIOSENSE WEBSTER INC | 31606488L | 10846835009804 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |