FDA Adverse Event
Other
Summary report: N
EZC HOSPTIAL BED W/HALF RAILS AND MATTRESS
MDR report key: 248983
·
Received November 9, 1999
Report
- Report Number
- 2182305-1999-00026
- Event Type
- Other
- Date Received
- November 9, 1999
- Date of Event
- October 1, 1999
- Report Date
- October 28, 1999
- Manufacturer
- SUNRISE MEDICAL CCG
- Product Code
- FNJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THAT A FEMALE RESIDENT WAS FOUND WITH HER HEAD CAUGHT BETWEEN THE 1/2 RAILS AND MATTRESS. RESIDENT SLIPPED OUT OF BED; THE SUNCORE MATTRESS SHE WAS ON COMPRESSED TO ABOUT 4", CREATING A GAP. RESIDENT WAS TAKEN TO THE E.R., EVALUATED AND RETURNED TO THE FACILITY, "BACK TO HER NORMAL SELF". PER SALES REP, HIS SOLUTION IS TO GO WITH THE IMPRINT MATTRESS FOR THIS FACILITY; IMPRINT IS AN EXTRA WIDE MATTRESS HIGH DENSITY FOAM ON THE EDGES WITH THE MIDDLE PROVIDING PRESSURE REDUCTION TO THE RESIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EZC HOSPTIAL BED W/HALF RAILS AND MATTRESS | MANUAL HOSPITAL BED W/ 1/2 RAILS & FOAM MATTRESS | FNJ | SUNRISE MEDICAL CCG | EZC BED, 1/2 RAILS, FOAM MATTR | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |