FDA Adverse Event Malfunction Summary report: N

ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT

MDR report key: 24887696 · Received April 15, 2026

Report

Report Number
3005094123-2026-00177
Event Type
Malfunction
Date Received
April 15, 2026
Date of Event
March 13, 2026
Report Date
April 15, 2026
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MMI
UDI-DI
00380740160630
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 8P13 (ALINITY I STAT HIGH SENSITIVE TROPONIN-I) THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 4Z21 (ALINITY I STAT HIGH SENSITIVITY TROPONIN-I) WITH 510K/PMA/BLA NUMBER K202525. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVITY TROPONIN-I RESULTS FOR A MALE PATIENT WHICH IS QUESTIONED BY THE DOCTOR. THE FOLLOWING DATA WAS PROVIDED (<26.2 = NORMAL): (B)(6) 2026 ALINITY RESULT = 86.4 UG/L, TROPONIN T RESULT = POSITIVE. HISTORICAL RESULTS: (B)(6) 2025 TROPONIN I = 55.4 CK = 136. (B)(6) 2024 TROPONIN I = 42.2 CK = 121 CKMB = 12. (B)(6) 2023 TROPONIN I = 106.1 CK = 130 CKMB = 8 MYOGLOBIN = 17.5 LDH = 139. (B)(6) 2023 TROPONIN I = 161.2 CK = 176 CKMB = 13 MYOGLOBIN = <7.51. (B)(6) 2022 TROPONIN I = 116.4 CK = 119 CKMB = 10 MYOGLOBIN = 32.7 LDH = 132. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
942806 ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ABBOTT IRELAND DIAGNOSTICS DIVISION 84471UD00 00380740160630

Patients

Seq Age Sex Outcome Treatment
1 NA Male ALNTY I PROCESSING MODU, 03R65-01, AI22530.