FDA Adverse Event Malfunction Summary report: N

NAVICARE NURSE CALL

MDR report key: 2487934 · Received March 6, 2012

Report

Report Number
2027454-2012-00021
Event Type
Malfunction
Date Received
March 6, 2012
Date of Event
February 29, 2012
Report Date
February 29, 2012
Manufacturer
HILL-ROM MANUFACTURING INC.
Product Code
ILQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION DETERMINED A TIMING ERROR MAY BE CREATED IF A ROOM AUDIO STATION RESETS INDEPENDENTS OF ITS ROOM CONTROL BOARD. A PROPER RESET SEQUENCE WILL CLEAR THE ERROR CONDITION. THE SOFTWARE WILL BE UPDATED TO HANDLE RESET EVENTS OF THIS NATURE SHOULD THEY OCCUR.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A BED EXIT CALL FOR ONE ROOM DID NOT SHOW ON THE NNC SYSTEM. ADDITIONAL TESTING IN THE ROOM FOUND DEVICES CONNECTED TO THE AUDIO STATION BED CONNECTOR WOULD NOT PLACE A CALL, NOR WOULD THE CODE LEVER ON THE ROOM AUDIO STATION. THE AUDIO STATION WOULD PLACE NORMAL AND STAFF EMERGENCY CALLS SUCCESSFULLY. THERE WAS NO IMPACT TO THE PT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NAVICARE NURSE CALL NAVICARE NURSE CALL ILQ HILL-ROM MANUFACTURING INC. P29000NNC NNC 3.0.200

Patients

Seq Age Sex Outcome Treatment
1