FDA Adverse Event
Malfunction
Summary report: N
NAVICARE NURSE CALL
MDR report key: 2487934
·
Received March 6, 2012
Report
- Report Number
- 2027454-2012-00021
- Event Type
- Malfunction
- Date Received
- March 6, 2012
- Date of Event
- February 29, 2012
- Report Date
- February 29, 2012
- Manufacturer
- HILL-ROM MANUFACTURING INC.
- Product Code
- ILQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION DETERMINED A TIMING ERROR MAY BE CREATED IF A ROOM AUDIO STATION RESETS INDEPENDENTS OF ITS ROOM CONTROL BOARD. A PROPER RESET SEQUENCE WILL CLEAR THE ERROR CONDITION. THE SOFTWARE WILL BE UPDATED TO HANDLE RESET EVENTS OF THIS NATURE SHOULD THEY OCCUR.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A BED EXIT CALL FOR ONE ROOM DID NOT SHOW ON THE NNC SYSTEM. ADDITIONAL TESTING IN THE ROOM FOUND DEVICES CONNECTED TO THE AUDIO STATION BED CONNECTOR WOULD NOT PLACE A CALL, NOR WOULD THE CODE LEVER ON THE ROOM AUDIO STATION. THE AUDIO STATION WOULD PLACE NORMAL AND STAFF EMERGENCY CALLS SUCCESSFULLY. THERE WAS NO IMPACT TO THE PT REPORTED.