FDA Adverse Event
Malfunction
Summary report: N
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
MDR report key: 24876330
·
Received April 14, 2026
Report
- Report Number
- 24876330
- Event Type
- Malfunction
- Date Received
- April 14, 2026
- Date of Event
- April 6, 2026
- Report Date
- April 7, 2026
- Manufacturer
- ICU MEDICAL
- Product Code
- CAI
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- PA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Description of Event or Problem · 0
THE BREATHING CIRCUITS FOR THE ANESTHESIA MACHINE WITH THE LOT NUMBER #6130212 HAVE A LEAK VOLUME >1000ML WHEN RUNNING A CIRCUIT TEST. THE PATIENT WAS IN THE OPERATING ROOM AND ANESTHESIA INDUCTION WAS DELAYED FOR 15 MINUTES WHILE TROUBLESHOOTING THE ANESTHESIA MACHINE. MULTIPLE COMPONENTS OF THE ANESTHESIA MACHINE WERE CHANGED BEFORE IDENTIFYING THAT THE CIRCUIT WAS THE ISSUE CAUSING THE LEAK. ANESTHESIA TECH MANAGER PRESENT IN OR TO TROUBLESHOOT MACHINE AND ENSURE MACH 406 WAS FUNCTIONING PROPERLY BEFORE INDUCING ANESTHESIA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 107675 | CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) | CAI | ICU MEDICAL | 6130212 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Male | Other |