FDA Adverse Event Malfunction Summary report: N

CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

MDR report key: 24876330 · Received April 14, 2026

Report

Report Number
24876330
Event Type
Malfunction
Date Received
April 14, 2026
Date of Event
April 6, 2026
Report Date
April 7, 2026
Manufacturer
ICU MEDICAL
Product Code
CAI
Product Problem
Yes
Report Source
User Facility report
Reporter Location
PA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

THE BREATHING CIRCUITS FOR THE ANESTHESIA MACHINE WITH THE LOT NUMBER #6130212 HAVE A LEAK VOLUME >1000ML WHEN RUNNING A CIRCUIT TEST. THE PATIENT WAS IN THE OPERATING ROOM AND ANESTHESIA INDUCTION WAS DELAYED FOR 15 MINUTES WHILE TROUBLESHOOTING THE ANESTHESIA MACHINE. MULTIPLE COMPONENTS OF THE ANESTHESIA MACHINE WERE CHANGED BEFORE IDENTIFYING THAT THE CIRCUIT WAS THE ISSUE CAUSING THE LEAK. ANESTHESIA TECH MANAGER PRESENT IN OR TO TROUBLESHOOT MACHINE AND ENSURE MACH 406 WAS FUNCTIONING PROPERLY BEFORE INDUCING ANESTHESIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
107675 CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) CAI ICU MEDICAL 6130212

Patients

Seq Age Sex Outcome Treatment
1 72 YR Male Other