BARRIGEL
Report
- Report Number
- 3014909464-2026-00034
- Event Type
- Injury
- Date Received
- April 14, 2026
- Date of Event
- March 19, 2026
- Report Date
- March 19, 2026
- Manufacturer
- PALETTE LIFE SCIENCES
- Product Code
- OVB
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). COMPLAINT VERIFICATION TESTING COULD NOT BE PERFORMED AS NO SAMPLE WAS AVAILABLE FOR ANALYSIS. ALL LOTS OF BARRIGEL ARE RELEASED BY BOTH GALDERMA/Q-MED (CONTRACT MANUFACTURER) AND PALETTE LIFE SCIENCES (LEGAL MANUFACTURER). BOTH RELEASES ENSURE THAT THE PRODUCT'S PREDETERMINED SPECIFICATIONS, AS NOTED IN THE RELEVANT PALETTE PART SPECIFICATIONS, ARE MET AND THAT THERE ARE NO CRITICAL DEVIATIONS ASSOCIATED WITH THE REVIEWED LOTS. WITHOUT THE DEVICE TO EVALUATE THE PROBABLE CAUSE COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. PALETTE LIFE SCIENCES WILL CONTINUE TO MONITOR AND TREND FOR REPORTS OF THIS NATURE.
IT WAS REPORTED THAT "SOME GEL MAY HAVE GOTTEN INTO THE VENOUS PLEXUS AND A LITTLE TO THE RIGHT INTERNAL ILIAC VEIN. FINDINGS CONSISTENT WITH INTRAVASATION OF SPACER GEL INTO THE PERIPROSTATIC VENOUS PLEXUS ON THE RIGHT SIDE AND EXTENSION INTO A RIGHT INTERNAL ILIAC VEIN TRIBUTARY ALONG THE PELVIC SIDEWALL." ADDITIONAL INFORMATION RECEIVED ON MARCH 24, 2026, INDICATED THAT "THE PHYSICIAN PERFORMED A FULL ANGIO WORKUP WITH CT AND CONFIRMED THERE WAS NO EXTENSION INTO THE COMMON ILIAC VEIN AND NO PULMONARY EMBOLUS. THE PATIENT IS ASYMPTOMATIC AND WAS PRESCRIBED 81 MG ASPIRIN WITH NO ADDITIONAL TREATMENT. THE PHYSICIAN WAS NOT INTERESTED IN SPEAKING WITH OUR FACULTY PHYSICIANS AS HE WOULDN'T CHANGE HIS TREATMENT PLAN BASED ON RESEARCH."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 940264 | BARRIGEL | HYDROGEL SPACER | OVB | PALETTE LIFE SCIENCES | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | NONE REPORTED. |