FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (HS-TNI ) CARTRIDGE

MDR report key: 24872879 · Received April 14, 2026

Report

Report Number
2245578-2026-00093
Event Type
Malfunction
Date Received
April 14, 2026
Date of Event
March 30, 2026
Report Date
May 20, 2026
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
PMA / PMN Number
K240984
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 0

ON 07-APR-2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT FALSE POSITIVE RESULT ON A 27-YEAR-OLD MALE PATIENT WITH CHEST PAIN. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED RESULT POS/NEG ALINITY (B)(6) 2026 1815 14 NEG I-STAT (B)(6) 2026 1834 31.5 POS ALINITY (B)(6) 2026 1925 15 NEG I-STAT (B)(6) 2026 1944 29.9 POS FACILITY POSITIVE CUT-OFF: I-STAT- 21. ALINITY <=27 NG/L. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY. 99TH PERCENTILE 90% CI SEX N (NG/L, PG/ML) (NG/L, PG/ML). FEMALE 490 13 (10, 17). MALE 404 28 (19, 58). OVERALL 895 21 (14, 30).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
159759 I-STAT TROPONIN (HS-TNI ) CARTRIDGE HS-TNI CARTRIDGE MMI ABBOTT POINT OF CARE INC. NA C26012A

Patients

Seq Age Sex Outcome Treatment
1 27 YR Male