I-STAT TROPONIN (HS-TNI ) CARTRIDGE
Report
- Report Number
- 2245578-2026-00093
- Event Type
- Malfunction
- Date Received
- April 14, 2026
- Date of Event
- March 30, 2026
- Report Date
- May 20, 2026
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- MMI
- PMA / PMN Number
- K240984
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON 07-APR-2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT FALSE POSITIVE RESULT ON A 27-YEAR-OLD MALE PATIENT WITH CHEST PAIN. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED RESULT POS/NEG ALINITY (B)(6) 2026 1815 14 NEG I-STAT (B)(6) 2026 1834 31.5 POS ALINITY (B)(6) 2026 1925 15 NEG I-STAT (B)(6) 2026 1944 29.9 POS FACILITY POSITIVE CUT-OFF: I-STAT- 21. ALINITY <=27 NG/L. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY. 99TH PERCENTILE 90% CI SEX N (NG/L, PG/ML) (NG/L, PG/ML). FEMALE 490 13 (10, 17). MALE 404 28 (19, 58). OVERALL 895 21 (14, 30).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 159759 | I-STAT TROPONIN (HS-TNI ) CARTRIDGE | HS-TNI CARTRIDGE | MMI | ABBOTT POINT OF CARE INC. | NA | C26012A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR | Male |