FDA Adverse Event Malfunction Summary report: N

INSYTE AUTOGUARD

MDR report key: 24863166 · Received April 13, 2026

Report

Report Number
1710034-2026-00406
Event Type
Malfunction
Date Received
April 13, 2026
Date of Event
March 18, 2026
Report Date
April 22, 2026
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825233
PMA / PMN Number
SEE H.11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: ROOT CAUSE COULDN'T BE DETERMINED DUE TO UNAVAILABILITY OF SAMPLE INFORMATION. A COMPLAINT HISTORY CHECK WAS PERFORMED, AND THIS IS THE 1ST RELATED COMPLAINT REPORTED WITH THE DEFECT/CONDITION OF NEEDLE RETRACTION FAILURE WITH LOT 5210685 REGARDING ITEM 382523. DHR NUMBER 5210685 HAS BEEN REVIEWED. NO RELATED QUALITY ISSUES OR PROCESS DEVIATIONS WERE FOUND DURING THE MANUFACTURE OF THE SUBASSEMBLY LOT AND PACKAGING LINE. A REVIEW OF THE APPLICABLE RISK DOCUMENTS INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY IN THE RISK MANAGEMENT DOCUMENTATION.

Additional Manufacturer Narrative · 0

THIS MDR CAPTURES EVENT 2 OF 3. G.4. K201075; K251654. H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE NEEDLE DID NOT RETRACT. IV NEEDLE DID NOT RETRACK WHEN WHITE BUTTON PRESSED TO ENGAGED SAFETY. ADDITIONAL INFORMATION 3/23/2026: PER STAFF: NO PATIENT HARM. THESE WERE NEAR MISS EVENTS THAT HAD THE POTENTIAL TO CAUSE INJURY TO EMPLOYEES. 4/6/2026 ADDITIONAL INFORMATION: DIFFERENT PATIENTS, SAME HCP. NO FURTHER INCIDENTS HAVE BEEN REPORTED TO ME AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
106431 INSYTE AUTOGUARD PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 5210685 00382903825233

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown