INSYTE AUTOGUARD
Report
- Report Number
- 1710034-2026-00406
- Event Type
- Malfunction
- Date Received
- April 13, 2026
- Date of Event
- March 18, 2026
- Report Date
- April 22, 2026
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FOZ
- UDI-DI
- 00382903825233
- PMA / PMN Number
- SEE H.11
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: ROOT CAUSE COULDN'T BE DETERMINED DUE TO UNAVAILABILITY OF SAMPLE INFORMATION. A COMPLAINT HISTORY CHECK WAS PERFORMED, AND THIS IS THE 1ST RELATED COMPLAINT REPORTED WITH THE DEFECT/CONDITION OF NEEDLE RETRACTION FAILURE WITH LOT 5210685 REGARDING ITEM 382523. DHR NUMBER 5210685 HAS BEEN REVIEWED. NO RELATED QUALITY ISSUES OR PROCESS DEVIATIONS WERE FOUND DURING THE MANUFACTURE OF THE SUBASSEMBLY LOT AND PACKAGING LINE. A REVIEW OF THE APPLICABLE RISK DOCUMENTS INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY IN THE RISK MANAGEMENT DOCUMENTATION.
THIS MDR CAPTURES EVENT 2 OF 3. G.4. K201075; K251654. H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.
NO ADDITIONAL INFORMATION.
IT WAS REPORTED THAT THE NEEDLE DID NOT RETRACT. IV NEEDLE DID NOT RETRACK WHEN WHITE BUTTON PRESSED TO ENGAGED SAFETY. ADDITIONAL INFORMATION 3/23/2026: PER STAFF: NO PATIENT HARM. THESE WERE NEAR MISS EVENTS THAT HAD THE POTENTIAL TO CAUSE INJURY TO EMPLOYEES. 4/6/2026 ADDITIONAL INFORMATION: DIFFERENT PATIENTS, SAME HCP. NO FURTHER INCIDENTS HAVE BEEN REPORTED TO ME AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 106431 | INSYTE AUTOGUARD | PERIPHERAL IV CATHETERS | FOZ | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 5210685 | 00382903825233 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |